MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2004.02.051

Медицинская библиотека · Journal of cataract and refractive surgery · 2004

Effect of a silicone intraocular lens with a sharp posterior optic edge on posterior capsule opacification.

Buehl Wolf, Menapace Rupert, Sacu Stefan, Kriechbaum Katharina, Koeppl Christina, Wirtitsch Matthias, Georgopoulos Michael, Findl Oliver

Клиническое исследованиеСравнительное исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the inhibiting effect on posterior capsule opacification (PCO) of a silicone intraocular lens (IOL) with a sharp posterior optic edge (ClariFlex OptiEdge, Advanced Medical Optics) and a silicone IOL with a round optic edge (PhacoFlex SI-40, Advanced Medical Optics).

SETTING

Department of Ophthalmology, University of Vienna, Vienna, Austria.

METHODS

This prospective randomized patient- and examiner-masked study comprised 104 eyes of 52 patients with bilateral age-related cataract. All patients had cataract surgery in both eyes and received a sharp-edged IOL in 1 eye and a round-edged IOL in the other eye. Postoperative examinations were at 1 week, 1 and 6 months, and 1 year. Digital slitlamp and retroillumination images were taken of each eye. The amount of PCO was assessed subjectively at the slitlamp and objectively using Automated Quantification of After-Cataract (AQUA) automated-image analysis software.

RESULTS

The sharp-edged IOL group had significantly less regeneratory and fibrotic PCO 1 month, 6 months, and 1 year after surgery. The mean AQUA PCO score (scale 0 to 10) was 0.71 in the sharp-edged IOL group and 1.40 in the round-edged IOL group (P<.001). The sharp-edged IOL group had less peripheral fibrotic PCO. There was no significant difference between the 2 IOL groups in patient reports of edge glare.

CONCLUSION

The sharp-edged design of the ClariFlex OptiEdge silicone IOL led to significantly less PCO than the round-edged PhacoFlex SI-40 IOL 1 year postoperatively.

Для цитирования: Buehl Wolf, Menapace Rupert, Sacu Stefan, Kriechbaum Katharina, Koeppl Christina, Wirtitsch Matthias, Georgopoulos Michael, Findl Oliver. Effect of a silicone intraocular lens with a sharp posterior optic edge on posterior capsule opacification.. Journal of cataract and refractive surgery. 2004. DOI: 10.1016/j.jcrs.2004.02.051.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.