Effect of haptic design on change in axial lens position after cataract surgery.

Wirtitsch Matthias G, Findl Oliver, Menapace Rupert, Kriechbaum Katharina, Koeppl Christina, Buehl Wolf, Drexler Wolfgang

Клиническое исследованиеСравнительное исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the postoperative changes in anterior chamber depth (ACD) between single-piece and multipiece AcrySof intraocular lenses (IOLs) (Alcon Laboratories, Inc.) and their effect on the postoperative refractive shift.

SETTING

Department of Ophthalmology, University of Vienna, Vienna, Austria.

METHODS

This prospective randomized bilateral study with intraindividual comparison comprised 104 eyes of 52 consecutive patients with age-related cataract. All patients had standardized cataract surgery with a temporal self-sealing incision, phacoemulsification, and IOL implantation in the capsular bag. The first 30 patients received a single-piece, 5.5 mm optic IOL (AcrySof SA30) in 1 eye and a multipiece IOL (AcrySof MA30) in the other eye. Another 22 patients received the corresponding 6.0 mm optic models (AcrySof SA60 and AcrySof MA60, respectively). Anterior chamber depth measurements and evaluation of the change in ACD during the first 6 postoperative months were performed using partial coherence laser interferometry. The amount of change in capsulorhexis size during the first 6 postoperative months was assessed using standardized retroillumination photographs.

RESULTS

Single-piece IOLs shifted significantly less postoperatively than multipiece IOLs. The multipiece IOLs shifted forward, especially from 1 day to 1 month. The change in the capsulorhexis area was similar between IOL types.

CONCLUSIONS

Single-piece AcrySof IOLs shifted minimally after the first postoperative day; therefore, the prescription for spectacles in eyes with this IOL can be obtained soon after surgery.

Для цитирования: Wirtitsch Matthias G, Findl Oliver, Menapace Rupert, Kriechbaum Katharina, Koeppl Christina, Buehl Wolf, Drexler Wolfgang. Effect of haptic design on change in axial lens position after cataract surgery.. Journal of cataract and refractive surgery. 2004. DOI: 10.1016/s0886-3350(03)00459-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.