MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/s0886-3350(03)00352-3

Медицинская библиотека · Journal of cataract and refractive surgery · 2003

Capsule measuring ring to predict capsular bag diameter and follow its course after foldable intraocular lens implantation.

Tehrani Mana, Dick H Burkhard, Krummenauer Frank, Pfirrmann Gudrun, Boyle Thomas, Stoffelns Bernhard M

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the extent of capsular bag shrinkage after cataract surgery with intraocular lens (IOL) implantation and develop a regression formula to predict postoperative capsular bag size.

SETTING

Eye Hospital, Johannes Gutenberg-University, Mainz, Germany.

METHODS

The axial length (AL), anterior chamber depth, and corneal radius in 58 eyes were measured preoperatively. Cataract surgery was by phacoemulsification followed by implantation of a 3-piece, acrylic, posterior chamber IOL. The capsular bag diameter and anterior capsulorhexis were measured intraoperatively and 1 day and 1, 3, and 6 months postoperatively using a Koch capsule measuring ring (HumanOptics).

RESULTS

The mean capsular bag size was 10.53 mm intraoperatively, 10.31 mm at 1 day, 9.62 mm at 1 month, 9.07 mm at 3 months, and 9.01 mm at 6 months. The mean capsular bag shrinkage over the entire postoperative period was 14.8% (P<.001). Of the parameters studied, only AL had a positive correlation with capsule shrinkage. The correlation was moderate but statistically significant (P =.001).

CONCLUSIONS

A correlation was found between capsular bag shrinkage and AL. Using preoperative biometric data, a regression formula of moderate validity was determined to predict capsular bag shrinkage.

Для цитирования: Tehrani Mana, Dick H Burkhard, Krummenauer Frank, Pfirrmann Gudrun, Boyle Thomas, Stoffelns Bernhard M. Capsule measuring ring to predict capsular bag diameter and follow its course after foldable intraocular lens implantation.. Journal of cataract and refractive surgery. 2003. DOI: 10.1016/s0886-3350(03)00352-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.