Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2009

Atrial fibrillation therapy in patients with a CRT defibrillator with wireless telemetry.

Crossley George H, Aonuma Kazutaka, Haffajee Charles, Shoda Morio, Meijer Albert, Bauer Alexander, Boriani Giuseppe, Svendsen Jesper, Thomas Shelley, Wiggenhorn Christopher, Unterberg-Buchwald Christina, Concerto-AT Study Investigators

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Atrial fibrillation (AF) is a major cause of morbidity and mortality, especially in patients with congestive heart failure.

OBJECTIVES

The purposes of this international, prospective multicenter study were to evaluate the efficacy of atrial shock therapy in patients with a cardiac resynchronization therapy defibrillator (CRT-D) and to evaluate the safety of the new CRT-D. The effectiveness of a new wireless telemetry system was also evaluated.

METHODS

A total of 282 patients, without permanent AF, who had indications for a CRT-D were included. Atrial shock therapy was tested on both spontaneous and induced AF episodes. The effectiveness of the Medtronic wireless telemetry system (Conexus; Medtronic Inc., Minneapolis, MN, USA) was also tested. Secondary endpoints included the heart failure Clinical Composite Response, system performance evaluation, and adverse event summary.

RESULTS

Atrial shock therapy was successful in 168 of 171 episodes (98.2%). Of these, 138 episodes were induced and 33 were spontaneous. Successful cardioversion occurred in 137 of the 138 induced-AF episodes (86.1% with 12 joule (J), 13.1% with 24 J, and 0.7% with 35 J). During the first 3 months of implant, there were 43 system-related complications in 37 subjects out of 278 subjects. There were 1,999 Conexus telemetry uses recorded during this study. This includes 282 uses during the implant procedure. There were no cases of complete loss of telemetry or any adverse events reported using this system.

CONCLUSION

We achieved an atrial shock efficacy of 98.2% in patients who met standard CRT-D indications. The wireless telemetry performed well with no reported unanticipated adverse device effects.

Для цитирования: Crossley George H, Aonuma Kazutaka, Haffajee Charles, Shoda Morio, Meijer Albert, Bauer Alexander, Boriani Giuseppe, Svendsen Jesper, Thomas Shelley, Wiggenhorn Christopher, Unterberg-Buchwald Christina, Concerto-AT Study Investigators. Atrial fibrillation therapy in patients with a CRT defibrillator with wireless telemetry.. Pacing and clinical electrophysiology : PACE. 2009. DOI: 10.1111/j.1540-8159.2009.02171.x.