DOI 10.1097/01.rvi.0000134497.50590.e2
Extended interval for retrieval of Günther Tulip filters.
Аннотация
PURPOSE: To evaluate the Gunther Tulip vena cava filter with regard to ease of placement, complications, and retrieval over long time periods.
MATERIALS AND
METHODS
In 53 patients (ratio of men to women, 24:29; mean age, 52.8 years) retrievable Gunther Tulip filters (Vena Cava M Reye Filter Set; William Cook Europe, Denmark) were inserted. Indications included planned major surgery with recent pulmonary embolus or high pulmonary embolus risk (n = 16), extensive ilio-femoral thrombus (n = 11), deep vein thrombosis with anticoagulant complications (n = 9), breakthrough pulmonary embolus despite anticoagulant therapy (n = 4), and contraindication to anticoagulant therapy (n = 13). All patients were followed-up for immediate and long-term complications.
RESULTS
Fifty-three filters were successfully placed in 52 of 53 patients, yielding a success rate of 98.1%. Nineteen patients underwent attempted retrieval of their filter. Sixteen of 19 retrieval procedures were successful (84%). In three patients, the filter could not be removed on attempted retrieval (extensive filter thrombus in two patients and attachment to the wall in one patient). One patient received two filters, which were both successfully retrieved at a later date. Median implantation time for retrievable filters was 34 days (range, 7-126 days). Mean follow-up for patients with permanent filters was 13 months. Two major complications (pneumothorax and break through pulmonary embolus) and three minor complications (right internal jugular vein thrombosis in two patients and transient Horner's Syndrome in one patient) were recorded.
CONCLUSION
Insertion and retrieval of filters is safe and feasible. Preliminary data suggest that Gunther Tulip filter retrieval is feasible over and above the manufacturer's recommended retrieval interval of 14 days.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.