Аннотация

BACKGROUND

Postoperative observed air leakage does not always originate from parenchymal defects but may arise from defects in the chest drainage unit, connections or reverse airflow in water seals. We investigated such false air leakage using a new chest drainage unit with a built-in CO2-detector and an electronic chest drainage unit.

METHODS

Two types of chest drainage units were tested in a simple porcine model: A well-known electronic chest drainage unit and a new chest drainage unit with integrated CO2-detector. We created a setup of true air leakage-a parenchymal lesion, and false air leakage-allowing air to flow into the thoracic cavity alongside the chest drain.

RESULTS

We demonstrated that the new chest drainage unit with a built-in CO2-detector can distinguish between experimentally induced true air leakage and false air leakage.

CONCLUSION

Available chest drainage systems do not allow direct assessment of true or false air leakage, which may increase chest drain duration unnecessarily. The integration of a CO2-sensitive color indicator into a chest drainage unit allows simple distinction between false air leak and true air leak, which may improve postoperative management.

Для цитирования: Eckardt Jens, Lijkendijk Marike, Licht Peter B, Stenger Michael. New Chest Drainage Unit with Integrated CO2-Detector Demonstrates False Air Leak.. Surgical innovation. 2021. DOI: 10.1177/15533506211047446.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.