Extracellular pH monitoring for use in closed-loop vagus nerve stimulation.

Cork Simon C, Eftekhar Amir, Mirza Khalid B, Zuliani Claudio, Nikolic Konstantin, Gardiner James V, Bloom Stephen R, Toumazou Christofer

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Vagal nerve stimulation (VNS) has shown potential benefits for obesity treatment; however, current devices lack physiological feedback, which limit their efficacy. Changes in extracellular pH (pHe) have shown to be correlated with neural activity, but have traditionally been measured with glass microelectrodes, which limit their in vivo applicability.

APPROACH: Iridium oxide has previously been shown to be sensitive to fluctuations in pH and is biocompatible. Iridium oxide microelectrodes were inserted into the subdiaphragmatic vagus nerve of anaesthetised rats. Introduction of the gut hormone cholecystokinin (CCK) or distension of the stomach was used to elicit vagal nerve activity.

MAIN

RESULTS

Iridium oxide microelectrodes have sufficient pH sensitivity to readily detect changes in pHe associated with both CCK and gastric distension. Furthermore, a custom-made Matlab script was able to use these changes in pHe to automatically trigger an implanted VNS device.

SIGNIFICANCE: This is the first study to show pHe changes in peripheral nerves in vivo. In addition, the demonstration that iridium oxide microelectrodes are sufficiently pH sensitive as to measure changes in pHe associated with physiological stimuli means they have the potential to be integrated into closed-loop neurostimulating devices.

Для цитирования: Cork Simon C, Eftekhar Amir, Mirza Khalid B, Zuliani Claudio, Nikolic Konstantin, Gardiner James V, Bloom Stephen R, Toumazou Christofer. Extracellular pH monitoring for use in closed-loop vagus nerve stimulation.. Journal of neural engineering. 2018. DOI: 10.1088/1741-2552/aa8239.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.