Clinical validation of an innovative dried whole-blood spot method to quantify simultaneously vancomycin and creatinine in adult patients.
СтатьяВалидационное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Hassanzai M, Bahmany S, van Onzenoort H A W, van Oldenrijk J, Koch B C P, de Winter B C M
Аннотация
BACKGROUND: A drawback of vancomycin use is the need for therapeutic drug monitoring and renal function monitoring. Traditional blood sampling involves drawing blood through a venepuncture. An alternative method, dried blood spot (DBS) sampling allows for self-sampling at home.
OBJECTIVES: To clinically validate a DBS method for simultaneous monitoring of vancomycin and creatinine.
METHODS: Hospitalized adults treated with intravenous vancomycin were included (trial registration NCT05257070). Blood sampling consisted of one venepuncture and one finger prick. Whole-blood DBS samples from patients were obtained by applying one drop of whole blood onto Whatman 903 filtrate paper. Bland-Altman analyses were used to assess the agreement and bias between the two measurements. Patients were asked to state their preferences for one of the two sampling methods.
RESULTS: The study involved a final analysis of 39 patient samples for the clinical validation of vancomycin and 46 patient samples for the clinical validation of creatinine. The difference between plasma and DBS concentrations was ≤20% for 77% of the vancomycin samples, the mean bias was -0.1379% (95% limit of agreement -5.899-5.623). The difference between plasma and DBS concentrations was ≤20% for 89% of the creatinine samples, the mean bias was 2.656% (95% limit of agreement -26.16-31.47). Most patients (18 out of 31) preferred a finger prick over a venepuncture and 12 patients indicated no preference.
CONCLUSIONS: This is the first study that successfully clinically validated a DBS sampling method for simultaneous measurement of vancomycin and creatinine, allowing for direct use in (outpatient) practice.
OBJECTIVES: To clinically validate a DBS method for simultaneous monitoring of vancomycin and creatinine.
METHODS: Hospitalized adults treated with intravenous vancomycin were included (trial registration NCT05257070). Blood sampling consisted of one venepuncture and one finger prick. Whole-blood DBS samples from patients were obtained by applying one drop of whole blood onto Whatman 903 filtrate paper. Bland-Altman analyses were used to assess the agreement and bias between the two measurements. Patients were asked to state their preferences for one of the two sampling methods.
RESULTS: The study involved a final analysis of 39 patient samples for the clinical validation of vancomycin and 46 patient samples for the clinical validation of creatinine. The difference between plasma and DBS concentrations was ≤20% for 77% of the vancomycin samples, the mean bias was -0.1379% (95% limit of agreement -5.899-5.623). The difference between plasma and DBS concentrations was ≤20% for 89% of the creatinine samples, the mean bias was 2.656% (95% limit of agreement -26.16-31.47). Most patients (18 out of 31) preferred a finger prick over a venepuncture and 12 patients indicated no preference.
CONCLUSIONS: This is the first study that successfully clinically validated a DBS sampling method for simultaneous measurement of vancomycin and creatinine, allowing for direct use in (outpatient) practice.