Proposal for fasting insulin and HOMA-IR reference intervals based on an extensive Brazilian laboratory database.
Аннотация
OBJECTIVE
Fasting insulin and the homeostasis model assessment of insulin resistance (HOMA-IR) index are relatively simple and reliable noninvasive markers of insulin resistance (IR). Given the relevance of correctly diagnosing IR, we emphasize the importance of establishing reliable reference intervals (RIs) for these markers. This study aimed to determine the RIs of fasting insulin and HOMA-IR index in adults living in Rio de Janeiro and, secondarily, to verify potential RI differences between sexes.
SUBJECTS AND
METHODS
Serum insulin levels of 146,497 individuals (ages 20-60 years) who underwent blood sampling in the state of Rio de Janeiro were obtained retrospectively through access to an extensive laboratory database. Insulin was measured using the electrochemiluminescence immunoassay method. After applying exclusion criteria, 21,684 individuals (18,576 [86%] women) were included. The RIs were estimated using a computational mining approach that integrates a selection of R packages.
RESULTS
The 95% RIs in women and men and in the overall population were, respectively, 2.54-13.30 μU/mL (15.3-80.12 pmol/L), 2.43-11.89 μU/mL (14.6-71.7 pmol/L), and 2.52-13.14 μU/mL (15.2-79.2 pmol/L) for fasting insulin levels and 0.39-2.86, 0.38-2.81, and 0.39-2.86 for HOMA-IR values. Despite significant differences in insulin levels and HOMA-IR index between men and women, the use of sex-specific RIs was not justified.
CONCLUSION
The RIs of fasting insulin levels and HOMA-IR values found in the overall population can be applied to both sexes. Thus, for our population, we suggest the RIs of 2.52-13.14 μU/mL (15.1-78.8 pmol/L) for fasting insulin and 0.39-2.86 for the HOMA-IR index.