DOI 10.1017/ice.2021.206

Медицинская библиотека · Infection control and hospital epidemiology · 2021

Efficacy of relatively low-cost ultraviolet-C light devices against Candida auris.

Pearlmutter Basya S, Haq Muhammed F, Cadnum Jennifer L, Jencson Annette L, Carlisle Matthew, Donskey Curtis J

СтатьяПоддержка службы здравоохранения СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Ultraviolet-C (UV-C) light devices could be useful to reduce environmental contamination with Candida auris. However, variable susceptibility of C. auris strains to UV-C has been reported, and the high cost of many devices limits their use in resource-limited settings.

OBJECTIVE

To evaluate the efficacy of relatively low-cost (<$15,000 purchase price) UV-C devices against C. auris strains from the 4 major phylogenetic clades.

METHODS

A modification of the American Society for Testing and Materials (ASTM) standard quantitative disk carrier test method (ASTM E 2197) was used to examine and compare the effectiveness of UV-C devices against C. auris, methicillin-resistant Staphylococcus aureus (MRSA), and bacteriophage Phi6. Reductions of 3 log10 were considered effective. UV-C irradiance measurements and colorimetric indicators were used to assess UV-C output.

RESULTS

Of 8 relatively low-cost UV-C devices, 6 met the criteria for effective decontamination of C. auris isolates from clades I and II, MRSA, and bacteriophage Phi6, including 3 room decontamination devices and 3 UV-C box devices. Candida auris isolates from clades III and IV were less susceptible to UV-C than clade I and II isolates; 1 relatively low-cost room decontamination device and 2 enclosed box devices met the criteria for effective decontamination of clade III and IV isolates. UV-C irradiance measurements and colorimetric indicator results were consistent with microorganism reductions.

CONCLUSIONS

Some relatively low-cost UV-C light technologies are effective against C. auris, including isolates from clades III and IV with reduced UV-C susceptibility. Studies are needed to evaluate the effectiveness of UV-C devices in clinical settings.

Для цитирования: Pearlmutter Basya S, Haq Muhammed F, Cadnum Jennifer L, Jencson Annette L, Carlisle Matthew, Donskey Curtis J. Efficacy of relatively low-cost ultraviolet-C light devices against Candida auris.. Infection control and hospital epidemiology. 2021. DOI: 10.1017/ice.2021.206.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.