m+Medelevator.RU
← Поиск публикаций
Heart rhythm · 2006

Direct measurements of membrane time constant during defibrillation strength shocks.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Sharma Vinod, Qu Fujian, Nikolski Vladimir P, DeGroot Paul, Efimov Igor R

Аннотация

BACKGROUND: Defibrillation shocks impose significant energy demand on implantable cardioverter-defibrillators (ICDs). Several modeling studies have been devoted to optimizing shock parameters, and a large number of these studies treat the heart as a simplified lumped network. The time constant of membrane polarization (tau(m)) is a key variable for such modeling efforts.

OBJECTIVE: The purpose of this study was to perform direct measurements of transmembrane potential (V(m)) during defibrillation strength shocks and estimate tau(m) of membrane polarization.

METHODS: A portion of the left ventricular epicardial surface of Langendorff-perfused rabbit hearts was stimulated using uniform electric fields produced by two parallel line electrodes. The V(m)s were recorded from di-4-ANEPPS-stained hearts using a multisite optical mapping system. The hearts were paced with 20 S1 pulses from the apex, and shocks (S2: 5, 10, 20 V/cm) were applied via the line electrodes during the action potential of the 20th S1 at two different coupling intervals (S1S2: 120 and 180 ms). Residual responses were obtained by subtracting responses to the 19th S1 from the responses to the 20th S1S2 pair and used for time-constant analysis by fitting a monoexponential function.

RESULTS: tau(m) exhibited a large variation and ranged from approximately 1 to 30 ms. Furthermore, tau(m) varied with electric field strength, S1S2 interval, position of the tissue from stimulating electrodes, and polarity of the response. To a large extent, the effects of all these factors were captured in a single parameter-the change in transmembrane voltage (DeltaV(m)) in response to the applied field (E). tau(m) showed a monotonically decreasing trend with DeltaV(m) for all Es and S1S2s.

CONCLUSION: Time constant of membrane polarization varies significantly during defibrillation strength shocks and shows a strong dependence on DeltaV(m).

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты