DOI 10.1111/j.1540-8159.2005.00229.x
Right heart failure due to loss of right ventricular capture in a patient with atrioventricular junction ablation and biventricular pacing.
Аннотация
We describe the case of a patient with atrioventricular (AV) junction ablation and chronic biventricular pacing in which intermittent dysfunction of the right ventricular (RV) lead resulted in left ventricular (LV) stimulation alone and onset of severe right heart failure. Restoration of biventricular pacing by increasing device output and then performing lead revision resolved the issue. This case provides evidence that LV pacing alone in patients with AV junction ablation may lead to severe right heart failure, most likely as a result of iatrogenic mechanical dyssynchrony within the RV. Thus, probably this pacing mode should be avoided in pacemaker-dependent patients with heart failure.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.