DOI 10.7860/jcdr/2015/13744.6828

Медицинская библиотека · Journal of clinical and diagnostic research : JCDR · 2015

Comparison of Root Surface Roughness Produced By Hand Instruments and Ultrasonic Scalers: An Invitro Study.

Kumar Pawan, Das Swarga Jyoti, Sonowal Saindhya Tora, Chawla Jitendra

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Instrumentation on tooth surface for debridement of hard and soft debris forms the basis of periodontal therapy. This involves periodic removal of accumulated material using different methods of instrumentation. An ideal instrument should eliminate all the deposits from the root surfaces with no or minimal alteration of the natural morphology.

AIM: To compare the root surface roughness after root planing performed with gracey curette and by ultrasonic scalers (Satelec P-5 Booster) set at different power modes.

MATERIALS AND

METHODS

The root surface roughness and its surface microtopography resulting from the use of Gracey curette, ultrasonic instrument at low, medium and high power setting on 35 healthy premolars extracted for orthodontic treatment purpose were examined using Optical Profilometer and the surface topography was assessed using Field Emission Microscope.

STATISTICAL ANALYSIS: Analysis of variance (ANOVA) test was used to observe the variance in a particular variable is partitioned into components attributable to different sources of variation. Duncan multiple range tests were used to determine whether three or more means differ significantly.

RESULTS AND

CONCLUSION

The mean roughness was found to be the highest in group where Scaling and Root Planing (SRP) was performed using ultrasonic scaler at low power mode (3.03±1.54 μm) whereas the lowest surface roughness was seen on the samples where SRP was performed using ultrasonic scaler at medium power mode. The surface roughness in group where SRP was performed with ultrasonic scaler at high power mode (2.22±0.74μm) was found to be similar to that of group in which root planing was carried out using curette (2.24±1.71μm).

Для цитирования: Kumar Pawan, Das Swarga Jyoti, Sonowal Saindhya Tora, Chawla Jitendra. Comparison of Root Surface Roughness Produced By Hand Instruments and Ultrasonic Scalers: An Invitro Study.. Journal of clinical and diagnostic research : JCDR. 2015. DOI: 10.7860/jcdr/2015/13744.6828.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.