Clinical Usefulness of the Active Fixation Quadripolar Left Ventricular Lead Compared With the Passive Fixation Quadripolar Lead in Cardiac Resynchronization Therapy.
Аннотация
BACKGROUND
This study compared the stability of the Medtronic Attain Stability Quad (ASQ), a novel quadripolar active fixation left ventricular (LV) lead with a side helix, to that of conventional quadripolar leads with passive fixation (non-ASQ) and evaluated their LV lead performance.
METHODS AND
RESULTS
In all, 183 consecutive patients (69 ASQ, 114 non-ASQ) who underwent cardiac resynchronization therapy (CRT) between January 2018 and June 2021 were enrolled. Complications, including elevated pacing capture threshold (PCT) levels, phrenic nerve stimulation (PNS), and LV lead dislodgement, were analyzed during the postimplantation period until the first outpatient visit after discharge. The frequency of LV lead-related complications was significantly lower in the ASQ than non-ASQ group (14% vs. 30%, respectively; P=0.019). Specifically, LV lead dislodgement occurred only in the non-ASQ group, and elevated PCT levels were significantly lower in the ASQ group (7% vs. 23%; P=0.007). Kaplan-Meier analysis confirmed a significantly lower incidence of LV lead-related complications in the ASQ group (log-rank P=0.005). Cox multivariable regression analysis showed a significant reduction in lead-related complications associated with ASQ (hazard ratio 0.44; 95% confidence interval 0.23-0.83; P=0.011).
CONCLUSIONS
The ASQ group exhibited fewer LV lead-related complications requiring reintervention and setting changes than the non-ASQ group. Thus, the ASQ may be a favorable choice for CRT device implantation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.