DOI 10.1016/j.hrthm.2022.11.015
Electrocardiogram Belt guidance for left ventricular lead placement and biventricular pacing optimization.
Аннотация
BACKGROUND
Patients with ischemic cardiomyopathy, non-left bundle branch block, or QRS duration <150 ms have a lower response rate to cardiac resynchronization therapy (CRT) than did other indicated patients. The ECG Belt system (EBS) is a novel surface mapping system designed to measure electrical dyssynchrony via the standard deviation of the activation times of the left ventricle.
OBJECTIVES
The objectives of this study were to evaluate the efficacy of the EBS in patients less likely to respond to CRT and to determine whether EBS use in lead placement guidance and device programming was superior to standard CRT care.
METHODS
This was a prospective randomized trial of patients with heart failure and EBS-guided CRT implantation and programming vs standard CRT care. The primary end point was relative change in left ventricular end-systolic volume from baseline to 6 months postimplantation.
RESULTS
A total of 408 patients from centers in Europe and North America were randomized. Although both patients with EBS and control patients had a mean improvement in left ventricular end-systolic volume, there was no significant difference in relative change from baseline (P = .26). While patients with a higher baseline standard deviation of the activation times derived greater left ventricular reverse remodeling, improvement in electrical dyssynchrony did not correlate with the extent of reverse remodeling.
CONCLUSION
The findings of the present study do not support EBS-guided therapy for CRT management of heart failure with reduced ejection fraction.