DOI 10.1016/j.cjca.2022.09.008
A Continuous Pacing and Recording Technique for Differentiating Left Bundle Branch Pacing From Left Ventricular Septal Pacing: Electrophysiologic Evidence From an Intrapatient-Controlled Study.
Аннотация
BACKGROUND
Left bundle branch pacing (LBBP) is a promising approach for achieving near-physiologic pacing. However, differentiating LBBP from left ventricular septal endocardial pacing (LVS(e)P) remains a challenge. This study aimed to establish a simple and effective method for differentiating LBBP from LVS(e)P and to evaluate their electrophysiologic characteristics.
METHODS
LBBP, using continuous uninterrupted pacing and real-time monitoring of electrocardiograms along with intracardiac electrograms, was performed in 97 consecutive patients. We evaluated the electrophysiologic characteristics observed during LBBP using 6 modalities: right ventricular septal pacing (RVSP), intraventricular septal pacing (IVSP 1 and 2), LVS(e)P, nonselective LBBP (NSLBBP), and selective LBBP (SLBBP).
RESULTS
Of the 97 patients, 87 (89.7%) met the criteria (abrupt change in paced QRS morphology with a transition from Qr to QR/qR in lead V1 and shortening of stimulus to V6 R-wave peak time [Stim-V6RWPT] of ≥ 10 ms with constant output while rather than after lead screwing) for nonselective left bundle branch (LBB) capture. Selective LBB capture was observed in 82 patients (84.5%). The Stim-V6RWPT of NSLBBP and SLBBP were significantly shorter than LVS(e)P (respectively, 67.1 ± 8.7 ms, 67.0 ± 9.3 ms, and 82.1 ± 10.9 ms). Stim-QRSend was the narrowest in IVSP2 (136.6 ± 15.2 ms) instead of NSLBBP (140.0 ± 17.1 ms).
CONCLUSIONS
The uninterrupted pacing technique for differentiating LBBP from LVS(e)P in the same group of patients is feasible. Electrophysiologic evidence from our intrapatient-controlled study shows that LBBP and LVS(e)P differ in ventricular electrical synchronization.