Dual-chamber ICD for left bundle branch area pacing: the cardiac resynchronization and arrhythmia sensing via the left bundle (cross-left) pilot study.

Clementy Nicolas, Bodin Alexandre, Ah-Fat Vincent, Babuty Dominique, Bisson Arnaud

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Left bundle branch area pacing (LBBAP) has emerged as a promising technique to deliver cardiac resynchronization therapy (CRT). However, safety and efficacy of ventricular arrhythmia sensing via the left bundle in implantable cardioverter-defibrillator (ICD) recipients remain unclear. We sought to evaluate the feasibility of a single LBBAP lead connected to a dual-chamber ICD in patients indicated with a CRT-D implantation.

METHODS

The CROSS-LEFT pilot study prospectively included 10 consecutive patients with a reduced ejection fraction and a complete left bundle branch block, indicated with a prophylactic CRT-D. A DF-1 lead was implanted at the right ventricular (RV) apex, and an LBBAP lead through the interventricular septum. Ventricular fibrillation was induced at implantation in both conventional (RV) and left bundle branch area sensing configurations. The latter was the final sensing configuration, and patients were implanted with a dual-chamber DF-1 ICD connected to the atrial lead (RA port), the LBBAP lead (RV IS-1 port), and the defibrillation lead (RV DF-1 port), the IS-1 pin being capped. Atrioventricular delay was optimized to ensure fusion between LBBAP and native conduction from the right bundle. Patients were followed during 6 months.

RESULTS

No difference between both configurations was observed regarding R-wave sensing in sinus rhythm (p = 0.22), ventricular fibrillation median interval detection (p = 1.00), or total induced episode duration (p = 0.78). LBBAP resulted in a significant reduction of median QRS width from 164 to 126 ms (p = 0.002). Median ventricular sensing significantly improved from 9.7 at implantation to 18.8 mV at 6 months (p = 0.01). Median LVEF also significantly improved from 29 to 44% at 6 months (p = 0.002).

CONCLUSION

Ventricular arrhythmia sensing and defibrillation can be performed via a single LBBAP lead connected to a dual-chamber ICD, and is associated with significant electromechanical reverse remodeling.

CLINICAL TRIAL REGISTRATION NUMBER: NCT05102227 In patients presenting with left bundle branch block and left ventricular systolic dysfunction, a left bundle branch area pacing lead connected to a DF-1 dual-chamber implantable cardioverter-defibrillator provides safe ventricular arrhythmia sensing and efficient electro-mechanical resynchronization.

Для цитирования: Clementy Nicolas, Bodin Alexandre, Ah-Fat Vincent, Babuty Dominique, Bisson Arnaud. Dual-chamber ICD for left bundle branch area pacing: the cardiac resynchronization and arrhythmia sensing via the left bundle (cross-left) pilot study.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2022. DOI: 10.1007/s10840-022-01342-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.