MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/aud.0000000000000158

Медицинская библиотека · Ear and hearing · 2015

Evaluating Inhibition of Motoneuron Firing From Electromyogram Data to Assess Vestibular Output Using Vestibular Evoked Myogenic Potentials.

Prakash S R, Herrmann Barbara S, Milojcic Rupprecht, Rauch Steven D, Guinan John J

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Vestibular evoked myogenic potentials (VEMPs) are due to vestibular responses producing brief inhibitions of muscle contractions that are detectable in electromyographic (EMG) responses. VEMP amplitudes are traditionally measured by the peak to peak amplitude of the averaged EMG response (VEMPpp) or by a normalized VEMPpp (nVEMPpp). However, a brief EMG inhibition does not satisfy the statistical assumptions for the average to be the optimal processing strategy. Here, it is postulated that the inhibition depth of motoneuron firing is the desired metric for showing the influence of the vestibular system on the muscle system. The authors present a metric called "VEMPid" that estimates this inhibition depth from the EMG data obtained in a usual VEMP data acquisition. The goal of this article was to compare how well VEMPid, VEMPpp, and nVEMPpp track inhibition depth.

DESIGN

To find a robust method to compare VEMPid, VEMPpp, and nVEMPpp, realistic physiological models for the inhibition of VEMP EMG signals were made using VEMP data from four measurement sessions on each of the five normal subjects. Each of the resulting 20 EMG-production models was adjusted to match the EMG autocorrelation of an individual subject and session. Simulated VEMP traces produced by these models were used to compare how well VEMPid, VEMPpp, and nVEMPpp tracked model inhibition depth.

RESULTS

Applied to simulated and real VEMP data, VEMPid showed good test-retest consistency and greater sensitivity at low stimulus levels than VEMPpp or nVEMPpp. For large-amplitude responses, nVEMPpp and VEMPid were equivalent in their consistency across subjects and sessions, but for low-amplitude responses, VEMPid was superior. Unnormalized VEMPpp was always worse than nVEMPpp or VEMPid.

CONCLUSIONS

VEMPid provides a more reliable measurement of vestibular function at low sound levels than the traditional nVEMPpp, without requiring a change in how VEMP tests are performed. The calculation method for VEMPid should be applicable whenever an ongoing muscle contraction is briefly inhibited by an external stimulus.

Для цитирования: Prakash S R, Herrmann Barbara S, Milojcic Rupprecht, Rauch Steven D, Guinan John J. Evaluating Inhibition of Motoneuron Firing From Electromyogram Data to Assess Vestibular Output Using Vestibular Evoked Myogenic Potentials.. Ear and hearing. 2015. DOI: 10.1097/aud.0000000000000158.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.