Intraobserver reproducibility and interobserver agreement of demarcation line depth measurements following corneal cross linking.

Tzamalis Argyrios, Vinciguerra Riccardo, Romano Vito, Arbabi Esmaeil, Batterbury Mark, Kaye Stephen B

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The aim of this study was to evaluate the intraobserver reproducibility of demarcation line depth measurement in keratoconic patients after epithelium-off corneal collagen cross linking and assess the interobserver variability/agreement among examiners of various experience levels.

METHODS

56 eyes of 56 patients undergoing corneal collagen cross linking for progressive keratoconus were enrolled. After 4 weeks, all patients underwent an anterior segment optical coherence tomography (AS-OCT; CASIA SS-1000; Tomey, Nagoya, Japan) in the cross-linked eye by four masked examiners (two medical students (S1 and S2) and two corneal fellows (F1 and F2)) in order to identify and measure the demarcation line depth in experienced and non-experienced users. The intraclass correlation coefficient and the coefficient of variation were calculated. Agreement between raters was evaluated using Bland-Altman plots, intraclass correlation coefficient and weighted-kappa statistics comparing raters in pairs.

RESULTS

The average demarcation line depth of all measurements was 254.34 ± 72.3 μm, varying from 84 to 459 μm. The intraclass correlation coefficient evaluating the reproducibility of measurements for F1 was 0.9379 (95% confidence interval (CI) =0.9035-0.9619), for the second (F2), it was 0.9837 (95% confidence interval = 0.9743-0.9901), while intraclass correlation coefficient between medical students was calculated 0.844. The overall intraclass correlation coefficient among all four observers was 0.8706 (95% confidence interval = 0.8061-0.9185). The coefficient of variation for repeated measurements was 5.981 μm (95% confidence interval = 3.966-7.471) and 3.312 μm (95% confidence interval = 2.468-3.981) for F1 and F2 raters, respectively. The percentage of demarcation line detection was 90.32%.

CONCLUSION

The reproducibility of demarcation line measures although very good (intraclass correlation coefficient > 0.9), yielded a difference between the two experienced raters. Furthermore, the novice raters did not reach an excellent level of agreement with the expert ones showing greater variability in their recordings.

Для цитирования: Tzamalis Argyrios, Vinciguerra Riccardo, Romano Vito, Arbabi Esmaeil, Batterbury Mark, Kaye Stephen B. Intraobserver reproducibility and interobserver agreement of demarcation line depth measurements following corneal cross linking.. European journal of ophthalmology. 2019. DOI: 10.1177/1120672119835116.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.