DOI 10.1186/s12938-019-0636-3

Медицинская библиотека · Biomedical engineering online · 2019

Novel dynamic corneal response parameters in a practice use: a critical review.

Jędzierowska Magdalena, Koprowski Robert

СтатьяОбзорПеревод пока не подготовлен

Аннотация

BACKGROUND

Non-contact tonometers based on the method using air puff and Scheimpflug's fast camera are one of the latest devices allowing the measurement of intraocular pressure and additional biomechanical parameters of the cornea. Biomechanical features significantly affect changes in intraocular pressure values, as well as their changes, may indicate the possibility of corneal ectasia. This work presents the latest and already known biomechanical parameters available in the new offered software. The authors focused on their practical application and the diagnostic credibility indicated in the literature.

DISCUSSION: An overview of available literature indicates the importance of new dynamic corneal parameters. The latest parameters developed on the basis of biomechanics analysis of corneal deformation process, available in non-contact tonometers using Scheimpflug's fast camera, are used in the evaluation of laser refractive surgery procedures, e.g. LASIK procedure. In addition, the assessment of changes in biomechanically corrected intraocular pressure confirms its independence from changes in the corneal biomechanics which may allow an intraocular pressure real assessment. The newly developed Corvis Biomechanical Index combined with the corneal tomography and topography assessment is an important aid in the classification of patients with keratoconus.

CONCLUSION

New parameters characterising corneal deformation, including Corvis Biomechanical Index and biomechanical compensated intraocular pressure, significantly extend the diagnostic capabilities of this device and may be helpful in assessing corneal diseases of the eye. Nevertheless, further research is needed to confirm their diagnostic pertinence.

Для цитирования: Jędzierowska Magdalena, Koprowski Robert. Novel dynamic corneal response parameters in a practice use: a critical review.. Biomedical engineering online. 2019. DOI: 10.1186/s12938-019-0636-3.

Предметные рубрики MeSH