Evaluation of corneal biomechanics in patients with keratectasia following LASIK using dynamic Scheimpflug analyzer.
Аннотация
PURPOSE: To investigate the corneal biomechanics in eyes with keratectasia following LASIK using a dynamic Scheimpflug analyzer.
DESIGN
Case-Control study.
METHOD: The subjects in the study included 12 eyes with keratectasia after LASIK (KE), 24 eyes with keratoconus (KC), 17 eyes without keratectasia after LASIK (LASIK), and 34 eyes with normal corneas (Normal). Corneal biomechanics of the four groups were evaluated using a dynamic Scheimpflug analyzer.
RESULTS
Compared with Normal (7.06 ± 0.54), the radius at the highest concavity (radius, mm) of LASIK (5.96 ± 0.76), KE (4.93 ± 0.61) and KC (5.39 ± 1.02) were significantly small. The Deflection Amplitude (HCDLA, mm) of Normal (0.94 ± 0.07) was significantly lower than those of KE (1.11 ± 0.10) and KC (1.06 ± 0.16), and was not significantly different from that of LASIK (0.98 ± 0.07). There were significant differences between LASIK and KE in radius and HCDLA (P < 0.05), whereas KE and KC had no differences in these parameters.
CONCLUSIONS
Corneal biomechanical features evaluated using the dynamic Scheimpflug analyzer suggest that biomechanical properties in eyes with keratectasia, keratoconus, and LASIK are different from those of normal eyes. Although the biomechanics in eyes with keratectasia differs from that in eyes with LASIK, it is similar to that in eyes with keratoconus.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.