MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/s25237338

Медицинская библиотека · Sensors (Basel, Switzerland) · 2025

Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications.

Maddaloni Francesca Saveria, Sarno Antonio, Loria Alessandro, Piai Anna, Lenardi Cristina, Esposito Antonio, Del Vecchio Antonella

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Photon-counting computed tomography (PCCT) represents a major technological advance in clinical CT imaging, offering superior spatial resolution, enhanced material discrimination, and potential radiation dose reduction compared to conventional energy-integrating detector systems. As the first clinically approved PCCT scanner becomes available, establishing a comprehensive characterization of its image quality is essential to understand its performance and clinical impact.

METHODS

Image quality was evaluated using a commercial quality assurance phantom with acquisition protocols typically used for three anatomical regions-head, abdomen/thorax, and inner ear-representing diverse clinical scenarios. Each region was scanned using both ultra-high-resolution (UHR, 120 × 0.2 mm slices) and conventional (144 × 0.4 mm slices) protocols. Conventional metrics, including signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), slice thickness accuracy, and uniformity, were assessed following international standards. Task-based analysis was also performed through target transfer function (TTF), noise power spectrum (NPS), and detectability index (d') to evaluate diagnostic relevance.

RESULTS

UHR protocols provided markedly improved spatial resolution, particularly in the inner ear imaging, as confirmed by TTF analysis, though with increased noise and reduced low-contrast detectability in certain conditions. CT numbers showed linear correspondence with known attenuation coefficients across all protocols.

CONCLUSIONS

This study establishes a detailed technical characterization of the first clinical PCCT scanner, demonstrating significant improvements in terms of spatial resolution and accuracy of the quantitative image analysis, while highlighting the need for noise-contrast optimization in high-resolution imaging.

Для цитирования: Maddaloni Francesca Saveria, Sarno Antonio, Loria Alessandro, Piai Anna, Lenardi Cristina, Esposito Antonio, Del Vecchio Antonella. Assessment of Image Quality Performance of a Photon-Counting Computed Tomography Scanner Approved for Whole-Body Clinical Applications.. Sensors (Basel, Switzerland). 2025. DOI: 10.3390/s25237338.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.