Effect of manual eyelid manipulation on intraocular pressure measurement by rebound tonometry.

Baek Sung Uk, Ha Ahnul, Kim Young Kook, Jeoung Jin Wook, Park Ki Ho

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND/AIMS: To investigate the effect of eyelid manipulation on the measurement of intraocular pressure (IOP) using two different tonometries (rebound tonometry (RT) vs Goldmann applanation tonometry (GAT)).

METHODS

103 patients with primary open-angle glaucoma were prospectively enrolled. For all of the patients, IOP measurements were performed in three different ways: (1) RT with lid manipulation (LM), (2) RT without LM and (3) GAT. The order of the three measurements was randomly selected. Additionally, the palpebral fissure height (PFH; elliptical space between upper and lower eyelids) was measured.

RESULTS

The mean value of IOP measured by GAT was 13.97±2.80 mm Hg, which was not significantly different from that by RT without LM (13.75±2.44 mm Hg; P=0.096), but which was significantly lower than that by RT with LM (15.21±2.91 mm Hg; P<0.001). On a Bland-Altman plot, RT with LM was overestimated relative to GAT (mean: -1.5) and RT without LM (mean: -1.2). Among the high IOPs (>20 mm Hg), interestingly, those measured by RT without LM were significantly lower than those measured by GAT (P<0.001). In the subgroup analysis of PFH, the smaller the PFH, the more exaggerated the IOP difference between GAT (P=0.014) and RT with LM (P<0.001).

CONCLUSION

RT-measured IOP was significantly exaggerated when manipulation was applied to the eyelid. This overall trend was more pronounced when PFH was small. GAT-measured IOP, meanwhile, showed a good correlation with IOP measured using RT without LM.

Для цитирования: Baek Sung Uk, Ha Ahnul, Kim Young Kook, Jeoung Jin Wook, Park Ki Ho. Effect of manual eyelid manipulation on intraocular pressure measurement by rebound tonometry.. The British journal of ophthalmology. 2018. DOI: 10.1136/bjophthalmol-2017-311587.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.