m+Medelevator.RU
← Поиск публикаций
Journal of cataract and refractive surgery · 2019

Rotational stability and refractive outcomes of a single-piece aspheric toric intraocular lens with 4 fenestrated haptics.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Savini Giacomo, Alessio Giovanni, Perone Giuseppe, Rossi Scipione, Schiano-Lomoriello Domenico

Аннотация

PURPOSE: To assess the outcomes of implantation of a single-piece toric intraocular lens (IOL) with 4 fenestrated haptics.

SETTING: IRCCS Fondazione Bietti, Rome, Italy.

DESIGN: Prospective case series.

METHODS: All patients who had implantation of the Mini Toric Ready IOL were consecutively enrolled. Intraoperatively, the IOL was aligned using an automated system. Follow-up visits were performed at 1 day, 1 week, and 1, 3, and 6 months. At each visit, retroillumination pictures were taken to assess IOL orientation; visual acuity and refraction were also measured.

RESULTS: The final analysis comprised 63 eyes (63 patients). From the first to the last follow-up, the mean arithmetic rotation was -0.2 degrees ± 3.5 (SD) (range -13 to +10 degrees) and the mean absolute rotation was 1.6 ± 3.1 degrees. Intraocular lens rotation from the first to the last examination was within 5 degrees in 92.1% of eyes and on consecutive visits, within 5 degrees in 98.4% or more of eyes. By 6 months, 10 IOLs (15.9%) had rotated clockwise and 10 counterclockwise. Linear regression did not show a statistically significant relationship between rotational stability and the axis of placement with any preoperative parameter (eg, axial length). The mean magnitude of preexisting corneal astigmatism was 1.9 ± 0.7 diopters (D) (range 0.76 to 3.72 D). At the last follow-up, the mean magnitude of refractive astigmatism was 0.5 ± 0.4 D (range 0.0 to 1.5 D); the difference was statistically significant (P < .05).

CONCLUSION: The toric IOL showed good rotational stability and is an option for correcting corneal astigmatism at the time of cataract surgery.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты