DOI 10.1016/j.jcrs.2019.03.020

Медицинская библиотека · Journal of cataract and refractive surgery · 2019

Comparative analysis of 2 swept-source optical coherence tomography biometers.

Sabatino Francesco, Matarazzo Francesco, Findl Oliver, Maurino Vincenzo

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To report the level of agreement, repeatability, and correlation of axial length (AL), mean keratometry, central corneal thickness, anterior chamber depth, lens thickness, and corneal diameter measurements of 2 swept-source optical coherence tomography (SS-OCT) biometers, the IOLMaster 700 (reference biometer) and the Argos (new biometer).

SETTING

Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.

DESIGN

Retrospective case series.

METHODS

Each patient had SS-OCT biometry with the reference biometer and new biometer. In addition to reporting the statistical results derived from both eyes, this study included a subgroup analysis of right eyes and left eyes. The level of agreement between the biometers was represented with the Bland-Altman method. Power vector analysis of the J0 (Jackson cross-cylinder, axes at 0 degrees and 90 degrees) and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) vectorial components of astigmatism was performed. Internal consistency was assessed with the Cronbach α coefficient of reliability. The dispersion of probability distribution was computed with the coefficient of variation. The intraoperator repeatability was calculated using the intraclass correlation coefficient (ICC).

RESULTS

The study comprised 112 patients (218 eyes). There was a statistically significant difference between the 2 biometers in all measurements (P < .05) except AL. The level of correlation between the reference biometer and the new biometer was very high for all the parameters except corneal diameter, and the agreement was high. The ICC and internal consistency were excellent with both biometers.

CONCLUSION

The new biometer provided good agreement and repeatability compared with the reference biometer.

Для цитирования: Sabatino Francesco, Matarazzo Francesco, Findl Oliver, Maurino Vincenzo. Comparative analysis of 2 swept-source optical coherence tomography biometers.. Journal of cataract and refractive surgery. 2019. DOI: 10.1016/j.jcrs.2019.03.020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.