Effects of decentration and tilt on the optical performance of 6 aspheric intraocular lens designs in a model eye.

Lawu Tjundewo, Mukai Koichiro, Matsushima Hiroyuki, Senoo Tadashi

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the effect of decentration and tilt on the optical performance of 6 aspheric intraocular lens (IOL) designs in a model eye.

SETTING

Department of Ophthalmology, Graduate School of Medicine, Dokkyo Medical University, Tochigi, Japan.

DESIGN

Experimental study.

METHODS

In theoretical simulations, the amount of spherical aberration in the IOL was varied to produce residual ocular spherical aberration (range -0.15 to 0.30 μm) at a 6.0 mm entrance pupil. Wavefront aberration analyses were performed with the ZEMAX optical design program (version August 20, 2014) to obtain the ocular root-mean-square values of astigmatism, coma, trefoil, and higher-order aberrations (HOAs) when the IOL was centered on the insertion position and misaligned at a 4.0 mm entrance pupil. The retinal visual images were calculated using the same conditions. Six 20.0 diopter (D) aspheric IOLs and one 20.0 D spherical IOL were used for the experimental studies. Each IOL was inserted in the model eye. The actual alignments were measured using a Scheimpflug camera (EAS-1000). The wavefront aberrations and visual images were gauged using a wavefront analyzer (KR-1W) at several IOL alignments.

RESULTS

Intraocular lens decentration and tilt increased wavefront aberrations and degraded optical performance. Astigmatism, coma, and HOAs generated by misaligned IOLs were related to the amount of spherical aberration correction of the IOLs. The extent of spherical aberration remained unchanged by the amount of misalignment. Experimental model eye results showed trends similar to theoretical results.

CONCLUSIONS

The spherical aberration correction amount in the aspheric IOL design was critical for the astigmatism, coma, and HOAs generated by the IOL misalignment. Additional spherical aberration corrections led to a more sensitive optical performance degradation resulting from IOL misalignment.

Для цитирования: Lawu Tjundewo, Mukai Koichiro, Matsushima Hiroyuki, Senoo Tadashi. Effects of decentration and tilt on the optical performance of 6 aspheric intraocular lens designs in a model eye.. Journal of cataract and refractive surgery. 2019. DOI: 10.1016/j.jcrs.2018.10.049.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.