Clinical ophthalmology (Auckland, N.Z.) · 2019
Long-term safety of posterior chamber implantable phakic contact lens for the correction of myopia.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sachdev Gitansha, Ramamurthy Dandapani
Аннотация
PURPOSE: To review the safety of the implantable phakic contact lens (IPCL V1, Caregroup Sight Solutions, India) for correction of myopia and myopic astigmatism.
METHODS: Retrospective interventional case series including 134 eyes which underwent IPCL implantation for correction of myopia and myopic astigmatism at a tertiary eye care hospital in South India. Intraoperative and postoperative complications were analyzed. The following parameters were analyzed at preoperative and 1, 6 and 12 month postoperative visits: visual acuity (in logMAR) and manifest refraction, slit lamp bio microscopy, dilated fundus examination, IOP and endothelial cell density measurement.
RESULTS: No intraoperative complications were noted. Cataract was the most common postoperative complication, wherein three eyes (2.2%) underwent observation for visually insignificant anterior subcapsular cataract and one eye (0.7%) required surgical intervention. Four eyes developed transient increased IOP due to steroid response (3 eyes) and pupillary block glaucoma (1 eye). The endothelial cell loss noted over a 1-year follow-up was 2.01%±4.12% (P=0.71). One eye developed hypopyon on the third postoperative day, which resolved subsequently with an unaided vision of 20/20. No vision threatening complications were noted. The mean follow-up in the cohort was 25.66±16.45 months.
CONCLUSION: The IPCL is a safe and effective treatment modality for correction of myopia and myopic astigmatism.
METHODS: Retrospective interventional case series including 134 eyes which underwent IPCL implantation for correction of myopia and myopic astigmatism at a tertiary eye care hospital in South India. Intraoperative and postoperative complications were analyzed. The following parameters were analyzed at preoperative and 1, 6 and 12 month postoperative visits: visual acuity (in logMAR) and manifest refraction, slit lamp bio microscopy, dilated fundus examination, IOP and endothelial cell density measurement.
RESULTS: No intraoperative complications were noted. Cataract was the most common postoperative complication, wherein three eyes (2.2%) underwent observation for visually insignificant anterior subcapsular cataract and one eye (0.7%) required surgical intervention. Four eyes developed transient increased IOP due to steroid response (3 eyes) and pupillary block glaucoma (1 eye). The endothelial cell loss noted over a 1-year follow-up was 2.01%±4.12% (P=0.71). One eye developed hypopyon on the third postoperative day, which resolved subsequently with an unaided vision of 20/20. No vision threatening complications were noted. The mean follow-up in the cohort was 25.66±16.45 months.
CONCLUSION: The IPCL is a safe and effective treatment modality for correction of myopia and myopic astigmatism.