Expert review of medical devices · 2018
Repeatability assessment of biometric measurements with different refractive states and age using a swept-source biometer.
СтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Martínez-Albert Noelia, Esteve-Taboada Jose J, Montés-Micó Robert, Fernández-Vega-Cueto Luis, Ferrer-Blasco Teresa
Аннотация
BACKGROUND: We aim to assess if refractive error or age may influence the repeatability of measurements using a swept-source biometer.
METHODS: A total of 61 subjects were evaluated with IOLMaster 700 acquiring measurements of axial length (AL), central corneal thickness (CCT), white-to-white distance (WTW), anterior aqueous depth (AQD), lens thickness (LT), and keratometry (K) readings. Repeatability was evaluated classifying the sample according to the refractive state and age by the calculation of the within-subject standard deviation (Sw), coefficient of repeatability, and intraclass correlation coefficient.
RESULTS: Very similar Sw was acquired among groups for AL, CCT, and WTW. Differences found for AQD, LT, keratometry and astigmatism among refractive and age groups were not clinically significant. Refractive groups obtained a Sw for AQD between 0.016 and 0.026, while for LT was between 0.023 and 0.029 mm. Besides, Sw showed values from 0.014 to 0.029 mm for keratometry. Age groups obtained a Sw of 0.025 and 0.016 mm for AQD, and 0.029 and 0.018 mm for LT, respectively. Sw was 0.019 vs 0.018 mm for K1, 0.014 vs 0.031 mm for K2.
CONCLUSIONS: IOLMaster 700 showed good repeatability for biometric and keratometric parameters in eyes classified according to their refractive error or age.
METHODS: A total of 61 subjects were evaluated with IOLMaster 700 acquiring measurements of axial length (AL), central corneal thickness (CCT), white-to-white distance (WTW), anterior aqueous depth (AQD), lens thickness (LT), and keratometry (K) readings. Repeatability was evaluated classifying the sample according to the refractive state and age by the calculation of the within-subject standard deviation (Sw), coefficient of repeatability, and intraclass correlation coefficient.
RESULTS: Very similar Sw was acquired among groups for AL, CCT, and WTW. Differences found for AQD, LT, keratometry and astigmatism among refractive and age groups were not clinically significant. Refractive groups obtained a Sw for AQD between 0.016 and 0.026, while for LT was between 0.023 and 0.029 mm. Besides, Sw showed values from 0.014 to 0.029 mm for keratometry. Age groups obtained a Sw of 0.025 and 0.016 mm for AQD, and 0.029 and 0.018 mm for LT, respectively. Sw was 0.019 vs 0.018 mm for K1, 0.014 vs 0.031 mm for K2.
CONCLUSIONS: IOLMaster 700 showed good repeatability for biometric and keratometric parameters in eyes classified according to their refractive error or age.