MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/13645706.2019.1599958

Медицинская библиотека · Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy · 2019

Modified implantation height of the Sapien3™ transcatheter heart valve.

Wendt Daniel, Shehada Sharaf-Eldin, König Luisa, Kahlert Philipp, Frey Ulrich, Mourad Fanar, Jakob Heinz, Thielmann Matthias, El Gabry Mohamed

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

Objectives

During transcatheter aortic valve implantation (TAVI), ideal positioning is crucial. The latest-generation balloon expandable Sapien3™ transcatheter heart valve (THV) comes with a marker, which is recommended to be exactly centered at the aortic annular level. We aimed to evaluate a higher "aortic" marker positioning.Material and methods: A total of 119 high-risk patients presenting with aortic stenosis were treated with the Sapien3™ THV. After having placed the THV more "aortic", clinical and hemodynamic data, especially postoperative pacemaker implantation and paravalvular leakages, were evaluated at 30-days according to VARC-2.

Results

The Sapien3™ THV was implanted in 92 patients via the transapical, in 13 patients via the transaortic and in 14 patients via the tranfemoral access. Mean age was 80.6 ± 5.7 years. Aortic valve area increased significantly (0.9 ± 0.3 vs. 1.80 ± 0.35cm2, p < .0001) and mean pressure gradients decreased from 41.0 ± 15.0 to 10.4 ± 3.5 mmHg (p < .0001). The majority of patients showed no or mild paravalvular aortic regurgitation (99.1%, 112/113), confirmed by transthoracic echocardiography at 30-days: PVL was absent or trace in 91.2% (103/113), mild in 7.9% (9/113) and moderate in 0.9% (1/113), whereas no patient developed severe PVL. Thirty days mortality was 5.0% (6/119). All patients (n = 113) were in NYHA functional class I or II at 30 days and three patients (2.5%) needed pacemaker implantation.

Conclusions

In conclusion, a modified higher "aortic" implantation of the Sapien3™ THV holds promise to further reduce paravalvular leakage as well as permanent pacemaker implantation in TAVI. This trial showed an extremely low postoperative pacemaker implantation rate of 2.5%.

Для цитирования: Wendt Daniel, Shehada Sharaf-Eldin, König Luisa, Kahlert Philipp, Frey Ulrich, Mourad Fanar, Jakob Heinz, Thielmann Matthias, El Gabry Mohamed. Modified implantation height of the Sapien3™ transcatheter heart valve.. Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy. 2019. DOI: 10.1080/13645706.2019.1599958.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.