MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jacc.2018.10.083

Медицинская библиотека · Journal of the American College of Cardiology · 2019

Durability of Transcatheter and Surgical Bioprosthetic Aortic Valves in Patients at Lower Surgical Risk.

Søndergaard Lars, Ihlemann Nikolaj, Capodanno Davide, Jørgensen Troels H, Nissen Henrik, Kjeldsen Bo Juel, Chang Yanping, Steinbrüchel Daniel Andreas, Olsen Peter Skov, Petronio Anna Sonia, Thyregod Hans Gustav Hørsted

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Transcatheter aortic valve replacement (TAVR) is an alternative to surgical aortic valve replacement (SAVR) in patients with severe aortic stenosis and intermediate or high surgical risk.

OBJECTIVES

The aim of this study was to compare the durability of transcatheter and surgical bioprosthetic aortic valves using standardized criteria.

METHODS

In the NOTION (Nordic Aortic Valve Intervention) trial, all-comer patients with severe aortic stenosis and lower surgical risk for mortality were randomized 1:1 to TAVR (n = 139) or SAVR (n = 135). Moderate/severe structural valve deterioration (SVD) was defined as a mean gradient ≥20 mm Hg, an increase in mean gradient ≥10 mm Hg from 3 months post-procedure, or more than mild intraprosthetic aortic regurgitation (AR) either new or worsening from 3 months post-procedure. Nonstructural valve deterioration (NSVD) was defined as moderate/severe patient-prosthesis mismatch at 3 months or moderate/severe paravalvular leakage. Bioprosthetic valve failure (BVF) was defined as: valve-related death, aortic valve reintervention, or severe hemodynamic SVD.

RESULTS

At 6 years, the rates of all-cause mortality were similar for TAVR (42.5%) and SAVR (37.7%) patients (p = 0.58). The rate of SVD was higher for SAVR than TAVR (24.0% vs. 4.8%; p < 0.001), whereas there were no differences in NSVD (57.8% vs. 54.0%; p = 0.52) or endocarditis (5.9% vs. 5.8%; p = 0.95). BVF rates were similar after SAVR and TAVR through 6 years (6.7% vs. 7.5%; p = 0.89).

CONCLUSIONS

In the NOTION trial through 6 years, SVD was significantly greater for SAVR than TAVR, whereas BVF was low and similar for both groups. Longer-term follow-up of randomized clinical trials will be necessary to confirm these findings. (Nordic Aortic Valve Intervention Trial; NCT01057173).

Для цитирования: Søndergaard Lars, Ihlemann Nikolaj, Capodanno Davide, Jørgensen Troels H, Nissen Henrik, Kjeldsen Bo Juel, Chang Yanping, Steinbrüchel Daniel Andreas, Olsen Peter Skov, Petronio Anna Sonia, Thyregod Hans Gustav Hørsted. Durability of Transcatheter and Surgical Bioprosthetic Aortic Valves in Patients at Lower Surgical Risk.. Journal of the American College of Cardiology. 2019. DOI: 10.1016/j.jacc.2018.10.083.