Immersion A-scan compared with partial coherence interferometry: outcomes analysis.

Packer Mark, Fine I Howard, Hoffman Richard S, Coffman Peggy G, Brown Laurie K

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare 2 methods of axial length measurement, immersion ultrasonography and partial coherence interferometry, and to elucidate surgical outcomes based on immersion measurements.

SETTING

Oregon Eye Institute, Eugene, Oregon, USA.

METHODS

Axial length measurements in 50 cataractous eyes were obtained by optical biometry (IOLMaster, Zeiss Humphrey Systems) and immersion ultrasound (Axis II, Quantel Medical), and the results were compared. Intraocular lens (IOL) power calculations in the same eyes after cataract extraction and posterior chamber IOL implantation were evaluated retrospectively based on the postoperative spherical equivalent prediction error.

RESULTS

Immersion ultrasonography and partial coherence interferometry measurements correlated in a highly positive manner (correlation coefficient = 0.996). Outcomes analysis demonstrated 92.0% of eyes were within +/-0.5 diopter of emmetropia based on immersion axial length measurements.

CONCLUSION

Immersion ultrasonography provided highly accurate axial length measurements and permitted highly accurate IOL power calculations.

Для цитирования: Packer Mark, Fine I Howard, Hoffman Richard S, Coffman Peggy G, Brown Laurie K. Immersion A-scan compared with partial coherence interferometry: outcomes analysis.. Journal of cataract and refractive surgery. 2002. DOI: 10.1016/s0886-3350(01)01259-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.