DOI 10.1016/s0886-3350(00)00790-2
Intraocular lens power calculation for lens exchange.
Аннотация
PURPOSE: To examine patients who had intraocular lens (IOL) exchange for large postoperative refractive errors and determine the factors that contributed to the error in IOL power calculation.
SETTING
Thirteen affiliated hospitals in Japan.
METHODS
This study comprised 34 cases that required IOL exchange because of large refractive errors after primary lens implantation. Patients with intraoperative complications were excluded from the study. The potential contribution of axial length, corneal refractive power, IOL manufacturer, and IOL fixation to errors in the predicted power was examined retrospectively. Axial length was calculated by the SRK/T and Holladay formulas using refraction after primary IOL implantation.
RESULTS
There was no statistical difference between the corneal refractive power before and after cataract surgery. The axial lengths calculated using the SRK/T and Holladay formulas were longer than the ultrasonic axial lengths in 24 and 23 cases, respectively. Using IOLs from the same manufacturer for both primary implantation and exchange reduced the error in predicted refraction.
CONCLUSION
Axial length and IOL manufacturer were important factors in predicting refraction power in eyes requiring IOL exchange.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.