Osmolality of various viscoelastic substances: comparative study.
Аннотация
PURPOSE: To determine the osmolality of various viscoelastic substances.
SETTING
Department of Ophthalmology, Johannes Gutenberg-University, Mainz, and Institute for Medical Device Testing, Memmingen, Germany.
METHODS
The analysis was carried out according to the European Pharmacopoeia by means of a calibrated osmometer using the freezing-point depression method. Each syringe was analyzed as a duplicate analysis.
RESULTS
Mean osmolalities (mOsmol/kg) of the sodium hyaluronate viscoelastic substances were Allervisc 299; Allervisc Plus 307; Amvisc Plus 335; AMO Vitrax 284; Biolon 279; Dispasan 311; Dispasan Plus 314; Healon 295; Healon GV 312; Healon5 322; HYA-Ophtal 376; Microvisc 313; Microvisc Plus 341; Provisc 307; Rayvisc 312; Viscoat 340; Visko 296; Visko Plus 319. Mean osmolarities of the hydroxypropyl methylcellulose viscoeslatic substances were Adatocel 278; HPMC Ophtal L 358; HPMC Ophtal H 360; La Gel 317; OcuCoat 309; Visco Shield 376.
CONCLUSIONS
There were significant differences in osmolality among viscoelastic substances, which may explain the differences in corneal thickness after cataract surgery. A viscoelastic substance with an osmolality of 305 mOsmol/kg or slightly higher is preferable, especially in patients with a compromised corneal endothelium.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.