DOI 10.1016/j.jcin.2023.03.035

Медицинская библиотека · JACC. Cardiovascular interventions · 2023

Bioprosthetic Valve Remodeling in Nonfracturable Surgical Valves: Impact on THV Expansion and Hydrodynamic Performance.

Meier David, Puehler Thomas, Lutter Georg, Shen Carol, Lai Althea, Gill Hacina, Akodad Mariama, Tzimas Georgios, Chhatriwalla Adnan, Allen Keith B, Blanke Philipp, Payne Geoffrey W, Wood David A, Leipsic Jonathon A, Webb John G, Sellers Stephanie L, Sathananthan Janarthanan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

There are limited data on the effect of bioprosthetic valve remodeling (BVR) on transcatheter heart valve (THV) expansion and function following valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) in a nonfracturable surgical heart valve (SHV).

OBJECTIVES

This study sought to assess the impact of BVR of nonfracturable SHVs on THVs after VIV implantation.

METHODS

VIV TAVR was performed using 23-mm SAPIEN3 (S3, Edwards Lifesciences) or 23/26-mm Evolut Pro (Medtronic) THVs implanted in 21/23-mm Trifecta (Abbott Structural Heart) and 21/23-mm Hancock (Medtronic) SHVs with BVR performed with a noncompliant TRUE balloon (Bard Peripheral Vascular Inc). Hydrodynamic assessment was performed, and multimodality imaging including micro-computed tomography was performed before and after BVR to assess THV and SHV expansion.

RESULTS

BVR resulted in limited improvement of THV expansion. The largest gain in expansion was observed for the S3 in the 21-mm Trifecta with up to a 12.7% increase in expansion at the outflow of the valve. Minimal change was observed at the level of the sewing ring. The Hancock was less amenable to BVR with lower final expansion dimensions than the Trifecta. BVR also resulted in notable surgical post flaring of up to 17.6°, which was generally more marked with the S3 than with the Evolut Pro. Finally, BVR resulted in very limited improvement in hydrodynamic function. Severe pinwheeling was observed with the S3, which improved slightly but persisted despite BVR.

CONCLUSIONS

When performing VIV TAVR inside a Trifecta and Hancock SHV, BVR had a limited impact on THV expansion and resulted in SHV post flaring with unknown consequences on coronary obstruction risk and long-term THV function.

Для цитирования: Meier David, Puehler Thomas, Lutter Georg, Shen Carol, Lai Althea, Gill Hacina, Akodad Mariama, Tzimas Georgios, Chhatriwalla Adnan, Allen Keith B, Blanke Philipp, Payne Geoffrey W, Wood David A, Leipsic Jonathon A, Webb John G, Sellers Stephanie L, Sathananthan Janarthanan. Bioprosthetic Valve Remodeling in Nonfracturable Surgical Valves: Impact on THV Expansion and Hydrodynamic Performance.. JACC. Cardiovascular interventions. 2023. DOI: 10.1016/j.jcin.2023.03.035.

Предметные рубрики MeSH