Discrepancy between invasive and echocardiographic transvalvular gradient after TAVI: Insights from the LAPLACE-TAVI registry.

Многоцентровое исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Yamazaki Chiemi, Higuchi Ryosuke, Saji Mike, Takamisawa Itaru, Nanasato Mamoru, Doi Shinichiro, Okazaki Shinya, Tamura Harutoshi, Sato Kei, Yokoyama Hiroaki, Onishi Takayuki, Tobaru Tetsuya, Shimizu Atsushi, Takanashi Shuichiro, Isobe Mitsuaki

Аннотация

BACKGROUND: Echocardiography-based transvalvular mean pressure gradient (ECHO-mPG) used to assess the forward valve function and structural valve deterioration could overestimate the true pressure gradient. This study evaluated the discrepancy between invasive and ECHO-mPG after transcatheter aortic valve implantation (TAVI) with respective valve type and size, its impact on a device success criterion, and predictors of a pressure discrepancy.

METHODS: We analyzed 645 patients registered in a multicenter TAVI registry (balloon-expandable valve [BEV]: 500; self-expandable valve [SEV]: 145). The invasive transvalvular mPG was measured after valve implantation using two Pigtail catheters (CATH-mPG), while the ECHO-mPG was measured within 48 h after TAVI. Pressure recovery (PR) was calculated using the following formula: ECHO-mPG × effective orifice area (EOA)/ascending aortic area (AoA) × (1 - EOA/AoA).

RESULTS: ECHO-mPG was weakly correlated with (r = 0.29, p < 0.0001), and consistently overestimated CATH-mPG in both BEV and SEV, and respective valve sizes. The magnitude of the discrepancy was larger for BEV than SEV (p < 0.001) and smaller valves (p < 0.001). After the correction of PR using the above formula, the pressure discrepancy remained for BEV (p < 0.001) but not SEV (p = 0.10). The proportion of patients with an ECHO-mPG > 20 mmHg decreased from 7.0% to 1.6% after correction (p < 0.0001). Among the baseline and procedural variables, post-procedural ejection fraction, BEV versus SEV, and smaller valves were associated with a larger discrepancy in mPG.

CONCLUSIONS: ECHO-mPG could be overestimated after TAVI, especially in patients with a smaller BEV. A higher ejection fraction, BEV, and smaller valves were predictors of a pressure discrepancy between CATH- and ECHO-mPG.