Digital Morphology Analyzer Sysmex DI-60 vs. Manual Counting for White Blood Cell Differentials in Leukopenic Samples: A Comparative Assessment of Risk and Turnaround Time.

Nam Minjeong, Yoon Sumi, Hur Mina, Lee Gun Hyuk, Kim Hanah, Park Mikyoung, Kim Hyeong Nyeon

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Digital morphology (DM) analyzers are increasingly being used for white blood cell (WBC) differentials. We assessed the laboratory efficiency of the Sysmex DI-60 system (DI-60; Sysmex, Kobe, Japan) in comparison with manual counting in leukopenic samples.

METHODS

In total, 40 peripheral blood smear samples were divided into normal, mild leukopenia, moderate leukopenia, and severe leukopenia groups based on WBC count. In each group, the risk and turnaround time (TAT) were compared between DI-60 and manual counting. Risk was determined by failure mode and effect analysis using the risk priority number (RPN) score, and TAT was recorded for the analytical phase.

RESULTS

Overall, DI-60 showed a five-fold lower risk (70 vs. 350 RPN) and longer TAT than manual counting. In severe leukopenic samples, DI-60 showed a shorter TAT/slide and a remarkably lower cell count/slide than manual counting. In all samples, the TAT/cell for DI-60 was substantially longer than that for manual counting (DI-60 vs. manual: total, 1.8 vs. 1.0 sec; normal, 1.5 vs. 0.7 sec; mild leukopenia, 1.9 vs. 0.9 sec; moderate leukopenia, 1.8 vs. 1.0 sec; severe leukopenia, 28.8 vs. 19.0 sec).

CONCLUSIONS

This is the first comparative assessment of risk and TAT between DI-60 and manual counting in leukopenic samples. DI-60 decreases the laboratory risk and improves patient safety, but requires more time to count fewer cells, especially in severe leukopenic samples. DM analyzers should be applied selectively depending on the WBC count to optimize laboratory efficiency.

Для цитирования: Nam Minjeong, Yoon Sumi, Hur Mina, Lee Gun Hyuk, Kim Hanah, Park Mikyoung, Kim Hyeong Nyeon. Digital Morphology Analyzer Sysmex DI-60 vs. Manual Counting for White Blood Cell Differentials in Leukopenic Samples: A Comparative Assessment of Risk and Turnaround Time.. Annals of laboratory medicine. 2022. DOI: 10.3343/alm.2022.42.4.398.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.