DOI 10.1136/bmjopen-2018-023775
Atrial fibrillation ablation with a spring sensor-irrigated contact force-sensing catheter compared with other ablation catheters: systematic literature review and meta-analysis.
Аннотация
OBJECTIVES
The objective of our review was to systematically assess available evidence on the effectiveness, safety and efficiency of a spring sensor-irrigated contact force (CF) catheter (THERMOCOOL SMARTTOUCH Catheter (ST)) for percutaneous ablation of paroxysmal or persistent atrial fibrillation (AF), compared with other ablation catheters, or with the ST with the operator blinded to CF data.
DESIGN
Systematic literature review and meta-analysis.
BACKGROUND
Emerging evidence suggests improved clinical outcomes of AF ablation using CF-sensing catheters; however, reviews to date have included data from multiple, distinct CF technologies.
METHODS
We conducted a systematic review and meta-analysis of published studies comparing the use of ST versus other ablation catheters for the treatment of AF. A comprehensive search of electronic and manual sources was conducted. The primary endpoint was freedom from recurrent atrial tachyarrhythmia (AT) at 12 months. Procedural and safety data were also analysed.
RESULTS
Thirty-four studies enrolling 5004 patients were eligible. The use of ST was associated with increased odds of freedom from AT at 12 months (71.0%vs60.8%; OR 1.454, 95% CI 1.12 to 1.88, p=0.004) over the comparator group, and the effect size was most evident in paroxysmal AF patients (75.6%vs64.7%; OR 1.560, 95% CI 1.09 to 2.24, p=0.015). Procedure and fluoroscopy times were shorter with ST (p=0.05 and p<0.01, respectively, vs comparator groups). The reduction in procedure time is estimated at 15.5 min (9.0%), and fluoroscopy time 4.8 min (18.7%). Complication rates, including cardiac tamponade, did not differ between groups.
CONCLUSIONS
Compared with the use of other catheters, AF ablation using the CF-sensing ST catheter for AF is associated with improved success rates, shorter procedure and fluoroscopy times and similar safety profile.