DOI 10.1111/jce.13472

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2018

Accuracy of left atrial bipolar voltages obtained by ConfiDENSE multielectrode mapping in patients with persistent atrial fibrillation.

Maille Baptiste, Das Moloy, Hussein Ahmed, Shaw Matthew, Chaturvedi Vivek, Morgan Maureen, Ronayne Christina, Snowdon Richard L, Gupta Dhiraj

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

The ConfiDENSE™ module (Carto3 v4) allows rapid annotation of endocardial electrograms acquired by multielectrode (ME) mapping. However, its accuracy in assessing atrial voltages is unknown.

METHODS AND

RESULTS

Two ConfiDENSE™ left atrial voltage maps were created during continuous pacing in 20 patients undergoing catheter ablation for persistent AF using a ME lasso catheter and a contact force (CF) sensing ablation catheter. The automated tissue proximity indicator (TPI) filter was then applied to the ME map to yield a TPI map. Reference maps (RM) were created offline by a blinded observer by manually assessing all points against fidelity criteria. Bipolar voltages and proportion of low voltage points (< 0.5 mV) derived from the ME, CF, and TPI maps were compared with those derived from the RM. Note that 853 ± 365 points, 252 ± 184 points, and 144 ± 73 were collected for ME, TPI, and CF maps, respectively, and 429 ± 153 points were included in the RM. Voltages with CF and TPI maps were similar to those with RM (1.57 ± 0.47 mV vs. 1.63 ± 0.31 mV, P = 0.57 and 1.50 ± 0.38 mV vs. 1.63 ± 0.31 mV, P = 0.07, respectively), whereas ME maps showed a significantly lower mean voltage (1.00 ± 0.22 mV, P < 0.001). As compared to RM maps (17 ± 8%), low voltage points were significantly overestimated by the ME maps (50 ± 9% (P < 0.001) and TPI maps (28 ± 13% (P < 0.001), but not by the CF maps (22 ± 14%, P = 0.17).

CONCLUSION

Application of the TPI filter to ConfiDENSE maps significantly increases the quality of the voltage data, conserving a reasonable point density, but still overestimates low voltage points as compared to CF-sensing maps or maps reviewed manually.

Для цитирования: Maille Baptiste, Das Moloy, Hussein Ahmed, Shaw Matthew, Chaturvedi Vivek, Morgan Maureen, Ronayne Christina, Snowdon Richard L, Gupta Dhiraj. Accuracy of left atrial bipolar voltages obtained by ConfiDENSE multielectrode mapping in patients with persistent atrial fibrillation.. Journal of cardiovascular electrophysiology. 2018. DOI: 10.1111/jce.13472.