DOI 10.1002/acm2.12618

Медицинская библиотека · Journal of applied clinical medical physics · 2019

Quantitative evaluation of breast density using a dual-energy technique on a digital breast tomosynthesis system.

Lu Kun-Mu, Yeh Da-Ming, Cao Bi-Hui, Lin Chia-Yi, Liang Chih-Yu, Zhou Yu-Bo, Tsai Chia-Jung

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Although breast density is considered a strong risk factor of breast cancer, its quantitative assessment is difficult. To investigate a quantitative method of measuring breast density using dual-energy mammographic imaging with central digital breast tomosynthesis in physically uniform and nonuniform phantoms.

MATERIAL AND

METHODS

The dual-energy imaging unit used a tungsten anode and silver filter with 30 kVp for high-energy images and 20 kVp for low-energy images. Uniform glandular-equivalent phantoms were used to calibrate a dual-energy based decomposition algorithm. The first study used uniform breast phantoms which ranged in thicknesses from 20 to 70 mm, in 10-mm increments, and which provided 30%, 50%, and 70% of breast density. The second study used uniform phantoms ranging from 10% to 90% of breast density. The third study used non-uniform phantoms (at an average density of 50%) with a thickness which ranged from 20 to 90 mm, in 10-mm increments.

RESULTS

The root mean square error of breast density measurements was 2.64-3.34% for the uniform, variable thickness phantoms, 4.17% for the uniform, variable density phantoms, and 4.49% for the nonuniform, variable thickness phantoms.

CONCLUSION

The dual-energy technique could be used to measure breast density with a margin of error of < 10% using digital breast tomosynthesis.

Для цитирования: Lu Kun-Mu, Yeh Da-Ming, Cao Bi-Hui, Lin Chia-Yi, Liang Chih-Yu, Zhou Yu-Bo, Tsai Chia-Jung. Quantitative evaluation of breast density using a dual-energy technique on a digital breast tomosynthesis system.. Journal of applied clinical medical physics. 2019. DOI: 10.1002/acm2.12618.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.