Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine · 2019
Diagnostic Performance of Automated Breast Ultrasound in Differentiating Benign and Malignant Breast Masses in Asymptomatic Women: A Comparison Study With Handheld Ultrasound.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Niu Lin, Bao Lingyun, Zhu Luoqian, Tan Yanjuan, Xu Xiaojing, Shan Yanna, Liu Jian, Zhu Qingqing, Jiang Chenxiang, Shen Yingzhao
Аннотация
OBJECTIVES: Our aim was to investigate the diagnostic potential of an automated breast ultrasound (ABUS) system in differentiating benign and malignant breast masses compared with handheld ultrasound (HHUS).
METHODS: Women were randomly and proportionally selected from outpatients and underwent both HHUS and ABUS examinations. Masses with final American College of Radiology Breast Imaging Reporting and Data System categories 2 and 3 were considered benign. Masses with final Breast Imaging Reporting and Data System categories 4 and 5 were considered malignant. The diagnosis was confirmed by pathologic results or at least a 1-year follow-up. Automated breast US and HHUS were compared on the basis of their sensitivity, specificity, positive predictive value, negative predictive value, and accuracy. Diagnostic consistency and areas under the receiver operating characteristic curves were analyzed. The maximum diameters of masses were compared among HHUS, ABUS, and pathologic results.
RESULTS: A total of 599 masses in 398 women were confirmed by pathologic results or at least a 1-year follow-up; 103 of 599 masses were malignant, and 496 were benign. There were no significant differences between ABUS and HHUS in terms of diagnostic accuracy (80.1% versus 80.6%), specificity (77.62% versus 80.24%), positive predictive value (46.12% versus 46.46%), and negative predictive value (97.96% versus 95.67%). There were significant differences in sensitivity (92.23% versus 82.52%; P < .01) and areas under the curve (0.85 versus 0.81; P < .05) between ABUS and HHUS. The correlation of the maximum diameter was slightly higher between ABUS and pathologic results (r = 0.885) than between HHUS and pathologic results (r = 0.855), but the difference was not significant (P > .05).
CONCLUSIONS: Automated breast US is better than HHUS in differentiating benign and malignant breast masses, especially with respect to specificity.
METHODS: Women were randomly and proportionally selected from outpatients and underwent both HHUS and ABUS examinations. Masses with final American College of Radiology Breast Imaging Reporting and Data System categories 2 and 3 were considered benign. Masses with final Breast Imaging Reporting and Data System categories 4 and 5 were considered malignant. The diagnosis was confirmed by pathologic results or at least a 1-year follow-up. Automated breast US and HHUS were compared on the basis of their sensitivity, specificity, positive predictive value, negative predictive value, and accuracy. Diagnostic consistency and areas under the receiver operating characteristic curves were analyzed. The maximum diameters of masses were compared among HHUS, ABUS, and pathologic results.
RESULTS: A total of 599 masses in 398 women were confirmed by pathologic results or at least a 1-year follow-up; 103 of 599 masses were malignant, and 496 were benign. There were no significant differences between ABUS and HHUS in terms of diagnostic accuracy (80.1% versus 80.6%), specificity (77.62% versus 80.24%), positive predictive value (46.12% versus 46.46%), and negative predictive value (97.96% versus 95.67%). There were significant differences in sensitivity (92.23% versus 82.52%; P < .01) and areas under the curve (0.85 versus 0.81; P < .05) between ABUS and HHUS. The correlation of the maximum diameter was slightly higher between ABUS and pathologic results (r = 0.885) than between HHUS and pathologic results (r = 0.855), but the difference was not significant (P > .05).
CONCLUSIONS: Automated breast US is better than HHUS in differentiating benign and malignant breast masses, especially with respect to specificity.