The prospective phase I "Flash-Skin I" trial: ultra-high dose rate radiotherapy implementation and quality assurance at a clinical linear accelerator.
СтатьяКлиническое исследование I фазыНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Dal Bello Riccardo, Psoroulas Serena, Flückiger Dominik, Krayenbühl Jerome, Moeckli Raphaël, Bailat Claude, Subiel Anna, Patallo Ileana Silvestre, von der Grün Jens, Balermpas Panagiotis, Guckenberger Matthias, Tanadini-Lang Stephanie
Аннотация
BACKGROUND AND PURPOSE: The combination of reduced normal tissue damage and unaffected tumor control is referred to as Flash-effect. This phenomenon has been observed with ultra-high dose rate (UHDR) radiotherapy in preclinical models. The Flash-Skin I (NCT06549439) treated seven patients with melanoma skin metastasis. The treatment protocol included 2 × 9 Gy UHDR followed by 1 × 9 Gy conventional radiotherapy. Here we report on the implementation of the study at a converted C-arm clinical linear accelerator.
MATERIALS AND METHODS: A dedicated preclinical radiation therapy quality assurance (RTQA) program was developed and the linac performance was validated for more than one year. Clinically, fourteen fractions of 9 Gy each were delivered with the 9 MeV UHDR electron linac. In-vivo measurements included: skin dose with radiochromic films, linac output with optically stimulated luminescent dosimeter (OSLD), pulse counting and inter-pulse stability with a scintillator, intra-pulse stability with a diamond detector.
RESULTS: The preclinical RTQA identified reduced linac output, which prompted beam tuning and additional quality assurance. This formed the basis for initiation of the clinical trial and successful treatment of all seven patients. All fourteen UHDR fractions showed a dose accuracy within 4.6% or better. The number of delivered pulses consistently matched the planned value, and film measurements confirmed the skin dose. Intra- and inter-pulse stability were within 3.8% and 2.8%, respectively.
CONCLUSION: The clinical implementation of Flash-Skin I at converted C-arm linac was successfully demonstrated and validated. The developed RTQA program may facilitate the development of future UHDR clinical trials.
MATERIALS AND METHODS: A dedicated preclinical radiation therapy quality assurance (RTQA) program was developed and the linac performance was validated for more than one year. Clinically, fourteen fractions of 9 Gy each were delivered with the 9 MeV UHDR electron linac. In-vivo measurements included: skin dose with radiochromic films, linac output with optically stimulated luminescent dosimeter (OSLD), pulse counting and inter-pulse stability with a scintillator, intra-pulse stability with a diamond detector.
RESULTS: The preclinical RTQA identified reduced linac output, which prompted beam tuning and additional quality assurance. This formed the basis for initiation of the clinical trial and successful treatment of all seven patients. All fourteen UHDR fractions showed a dose accuracy within 4.6% or better. The number of delivered pulses consistently matched the planned value, and film measurements confirmed the skin dose. Intra- and inter-pulse stability were within 3.8% and 2.8%, respectively.
CONCLUSION: The clinical implementation of Flash-Skin I at converted C-arm linac was successfully demonstrated and validated. The developed RTQA program may facilitate the development of future UHDR clinical trials.