Schirmer strip wetting length varies by commercial manufacturer.
Аннотация
OBJECTIVE
To determine whether wetting length of Schirmer tear test (STT) strips varies by commercial manufacturer of test.
ANIMALS STUDIED: Ten normal female beagle cross-breed dogs.
PROCEDURES: Schirmer tear test strips from four commercial manufacturers were used to assess wetting lengths in-vitro when exposed to a set volume of water over 1 minute. Digital photography was used to assess the total surface area of Schirmer strips from each manufacturer. Schirmer tear test type 1 was performed on normal dogs over 1 minute with STT strips from the same four commercial manufacturers in a randomized crossover design. ANOVA was used to detect differences between strips from different manufacturers.
RESULTS
Significant differences in wetting length were found between STT strips from different manufacturers in-vitro (P < .0001) and in-vivo (P < .0001). STT wetting lengths for the in-vitro experiments (mean ± SD) were 21.2 ± 0.8 mm (Amcon Laboratories), 27.8 ± 2.7 mm (Merck Animal Health), 30.0 ± 1.7 mm (HUB pharmaceuticals), and 31.5 ± 2.3 mm (Gulden Ophthalmics). STT wetting lengths for the experiments in live dogs (mean ± SD) were 17.5 ± 3.4 mm (Amcon Laboratories), 24.8 ± 3.9 mm (Merck Animal Health), 26.5 ± 3.7 mm (HUB pharmaceuticals), and 24.0 ± 3.9 mm (Gulden Ophthalmics). The surface area of STT strips were found to vary between different manufacturers.
CONCLUSIONS
Schirmer tear test type 1 results in dogs are affected by choice of STT strip commercial manufacturer. It is proposed that guidelines are created for the standardization of future STT strip production.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.