Short-term effects of humidification devices on respiratory pattern and arterial blood gases during noninvasive ventilation.
Аннотация
BACKGROUND
The impact of humidification devices on ventilatory and arterial blood gases parameters during noninvasive ventilation (NIV) remains controversial. The aim of the study was to compare the short-term impact of heat and moisture exchangers (HMEs) and heated humidifiers (HHs) during NIV for either hypercapnic or hypoxemic acute respiratory failure.
METHODS
Consecutive subjects receiving NIV were successively treated with HME and HH in randomized order for 30 min each. At the end of each period, arterial blood gases were measured and ventilatory parameters were recorded.
RESULTS
Eighty-one subjects were enrolled, of whom 52 were hypercapnic (with or without acidosis) and 29 hypoxemic. Minute ventilation was greater with the HME, in comparison with the HH (15 [12-18] vs 12 [10-16] median [interquartile range], P < .001), while P(aCO(2)) was increased when using HME, indicating a dead space effect. This effect was observed in all subjects, but was more pronounced in hypercapnic subjects (P(aCO(2)) 62 ± 17 mm Hg with HME vs 57 ± 14 with HH, P < .001). In a subgroup of 19 subjects with respiratory acidosis, alveolar hypoventilation improved only with the HH. The amplitude of the dead space impact was a function of the degree of hypercapnia.
CONCLUSIONS
Use of an HME decreased CO(2) elimination during NIV, despite increased minute ventilation, especially in hypercapnic subjects.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.