The Journal of clinical endocrinology and metabolism · 2020

Diagnosis of Primary Aldosteronism by Seated Saline Suppression Test-Variability Between Immunoassay and HPLC-MS/MS.

СтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАВалидационное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Thuzar Moe, Young Karen, Ahmed Ashraf H, Ward Greg, Wolley Martin, Guo Zeng, Gordon Richard D, McWhinney Brett C, Ungerer Jacobus P, Stowasser Michael

Аннотация

BACKGROUND: In primary aldosteronism (PA), excessive, autonomous secretion of aldosterone is not suppressed by salt loading or fludrocortisone. For seated saline suppression testing (SSST), the recommended diagnostic cutoff 4-hour plasma aldosterone concentration (PAC) measured by high-performance liquid chromatography-mass spectrometry (HPLC-MS/MS is 162 pmol/L. Most diagnostic laboratories, however, use immunoassays to measure PAC. The cutoff for SSST using immunoassay is not known. We hypothesized that the cutoff is different between the assays.

METHODS: We analyzed 80 of the 87 SSST tests that were performed during our recent study defining the HPLC-MS/MS cutoff. PA was confirmed in 65 by positive fludrocortisone suppression testing (FST) and/or lateralization on adrenal venous sampling and excluded in 15 by negative FST. PAC was measured by a chemiluminescence immunoassay (PACIA) in the SSST samples using the DiaSorin Liaison XL analyzer, and receiver operating characteristics (ROC) analysis was performed to identify the PACIA cutoff.

RESULTS: ROC revealed good performance (area under the curve = 0.893; P < .001) of 4-hour postsaline PACIA for diagnosis of PA and an optimal diagnostic cutoff of 171 pmol/L, with sensitivity and specificity of 95.4% and 80.0%, respectively. A higher cutoff of 217 pmol/L improved specificity (86.7%) with lower sensitivity (86.2%). PACIA measurements strongly correlated with PAC measured by HPLC-MS (r = 0.94, P < .001).

CONCLUSIONS: A higher diagnostic cutoff for SSST should be employed when PAC is measured by immunoassay rather than HPLC-MS/MS. The results suggest that (i) PA can be excluded if 4-hour PACIA is less than 171 pmol/L, and (ii) PA is highly likely if the PACIA is greater than 217 pmol/L by chemiluminescence immunoassay. A gray zone exists between the cutoffs of 171 and 217 pmol/L, likely reflecting a lower specificity of immunoassay.