Quantification of stenotic mitral valve area and diagnostic accuracy of mitral stenosis by dual-source computed tomography in patients with atrial fibrillation: comparison with cardiovascular magnetic resonance and transthoracic echocardiography.

Сравнительное исследованиеОценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kim Song Soo, Ko Sung Min, Song Meong Gun, Chee Hyun Kun, Kim Jun Suk, Hwang Hweung Kon, Lee Jae-Hwan

Аннотация

This study aimed to evaluate the utility of dual-source computed tomography (DSCT) for quantification of the mitral valve area (MVA) in patients with atrial fibrillation (AF) and mitral stenosis (MS) and to compare the results of DSCT with those of cardiovascular magnetic resonance (CMR) and transthoracic echocardiography (TTE). One hundred-two patients with AF and MS who had undergone electrocardiography-gated DSCT, TTE and CMR prior to operation were retrospectively enrolled. The MVA was planimetrically determined by DSCT, CMR, and TTE, as well as by Doppler TTE using the pressure half-time method (TTE-PHT). Agreement, relationship between measurements, and the highest accuracy were evaluated using Bland-Altman, Pearson correlation, and receiver operating characteristic analyses. The MVA on DSCT (mean, 1.27 ± 0.27 cm(2)) was significantly larger than that on CMR (1.15 ± 0.28 cm(2), P < 0.05), TTE-planimetry and TTE-PHT (1.16 ± 0.28 and 1.07 ± 0.30 cm(2), respectively; P < 0.05). TTE-planimetry had better correlation with planimetry on DSCT and CMR (r = 0.65 and 0.67, respectively; P < 0.05) than TTE-PHT (r = 0.51 and 0.55, respectively; P < 0.05). Using an MVA of 1.0 cm(2) on TTE-planimetry and TTE-PHT as the reference, the optimal thresholds for detecting severe MS on DSCT was 1.19 cm(2). The planimetry of the MVA measured by DSCT may be a reliable, alternative method for the quantification of MS in patients with AF.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты