Improved High-Frequency Ultrasound Corneal Biometric Accuracy by Micrometer-Resolution Acoustic-Property Maps of the Cornea.

Rohrbach Daniel, Silverman Ronald H, Chun Dan, Lloyd Harriet O, Urs Raksha, Mamou Jonathan

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Mapping of epithelial thickness (ET) is useful for detection of keratoconus, a disease characterized by corneal thinning and bulging in which epithelial thinning occurs over the apex. In prior clinical studies, optical coherence tomography (OCT) measurements of ET were systematically thinner than those obtained by 40-MHz high-frequency ultrasound (HFU) where a constant speed of sound (c) of 1636 m/s was used for all corneal layers. The purpose of this work was to study the acoustic properties, that is, c, acoustic impedance (Z), and attenuation (α) of the corneal epithelium and stroma independently using a scanning acoustic microscope (SAM) to investigate the discrepancy between OCT and HFU estimates of ET.

METHODS

Twelve unfixed pig corneas were snap-frozen and 6-μm sections were scanned using a custom-built SAM with an F-1.08, 500-MHz transducer and a 264-MHz bandwidth. Two-dimensional maps of c, Z, and α with a spatial resolution of 4 μm were derived.

RESULTS

SAM showed that the value of c in the epithelium (i.e., 1548 ± 18 m/s) is substantially lower than the value of c in the stroma (i.e., 1686 ± 33 m/s).

CONCLUSION

SAM results demonstrated that the assumption of a constant value of c for all corneal layers is incorrect and explains the prior discrepancy between OCT and HFU ET determinations.

TRANSLATIONAL RELEVANCE: The findings of this study have important implications for HFU-based ET measurements and will improve future keratoconus diagnosis by providing more-accurate ET estimates.

Для цитирования: Rohrbach Daniel, Silverman Ronald H, Chun Dan, Lloyd Harriet O, Urs Raksha, Mamou Jonathan. Improved High-Frequency Ultrasound Corneal Biometric Accuracy by Micrometer-Resolution Acoustic-Property Maps of the Cornea.. Translational vision science & technology. 2018. DOI: 10.1167/tvst.7.2.21.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.