DOI 10.1097/ico.0000000000001373
Evaluation of the Reliability and Repeatability of Scheimpflug System Measurement in Keratoconus.
Аннотация
PURPOSE: To assess the reliability and repeatability of the parameters obtained using a Scheimpflug system (Pentacam HR-high resolution) for monitoring progression in eyes with keratoconus.
METHODS
A prospective study was conducted based on measurements taken with the Pentacam HR in 82 eyes with keratoconus. We calculated the intraclass correlation coefficient (ICC) and the repeatability limit for topographic and aberrometric parameters.
RESULTS
All the parameters analyzed had excellent ICCs, above 0.95 in all cases, except for the measurement of the anterior best-fit sphere, for which the ICC was 0.873. The key repeatability limits for the parameters analyzed were as follows: 1.54 diopters (D) for the maximum keratometry in the central cornea, 1.08 D for maximum keratometry, 32.6 degrees for the axis of corneal astigmatism, 11.9 degrees for the coma axis, and 14.2 μm for thinnest corneal thickness.
CONCLUSIONS
The level of repeatability of the measurements obtained with a Pentacam HR confirms the usefulness of this system for the diagnosis of keratoconus. Nevertheless, it has certain limitations for monitoring individual progression of keratoconus and planning intrastromal corneal ring segment insertion.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.