DOI 10.3928/1081597x-20170621-03
Comparison of Corneal Deformation Parameters in Keratoconic and Normal Eyes Using a Non-contact Tonometer With a Dynamic Ultra-High-Speed Scheimpflug Camera.
Аннотация
PURPOSE: To evaluate and compare biomechanical properties in normal and keratoconic eyes using a dynamic ultra-high-speed Scheimpflug camera equipped with a non-contact tonometer (Corvis ST; Oculus Optikgeräte GmbH, Wetzlar, Germany).
METHODS
This retrospective study evaluated 89 eyes (47 normal, 42 keratoconic) and a validation arm of 72 eyes (33 normal, 39 keratoconic) using the Corvis ST. A diagnosis of keratoconus was established by clinical findings confirmed by topography and tomography. Dynamic corneal response parameters collected by the Corvis ST (A1 velocity, deformation amplitude [DA], DA Ratio Max 1mm, and Max Inverse Radius) and a stiffness parameter at first applanation (SP-A1) were incorporated into a novel logistic regression equation (DCR index). Area under the receiver operating curve (AUC) was used to assess the sensitivity and specificity of the DCR index.
RESULTS
DA, DA Ratio Max 1mm, Max Inverse Radius, and SP-A1 were each found to be statistically significantly different between normal and keratoconic eyes (Mann-Whitney test [independent samples]; P = .0077, < .0001, < .0001, and < .0001, respectively; significance level: P < .05). DCR index demonstrated high sensitivity, specificity, and overall correct detection rate (92.9%, 95.7%, and 94.4%, respectively; AUC = 98.5). The sensitivity and overall correct detection rate improved when eyes with Topographical Keratoconus Classification grades (TKC) greater than 0 were reevaluated (from 92.9% to 96.6% and from 94.4% to 96.1%, respectively).
CONCLUSIONS
Combining multiple biomechanical parameters (A1 velocity, DA, DA Ratio Max 1mm, Max Inverse Radius, and SP-A1) into a logistic regression equation allows for high sensitivity and specificity for distinguishing keratoconic from normal eyes. [J Refract Surg. 2017;33(9):625-631.].