DOI 10.1097/opx.0000000000001120
Tonometry after Intrastromal Corneal Ring Segments for Keratoconus.
Аннотация
SIGNIFICANCE: Reliable intraocular pressure (IOP) measurement after intrastromal corneal ring segments (ICRS) implantation is a challenge because of altered corneal morphology. In this study, IOP is measured with four tonometers, compared with Goldmann applanation tonometry (GAT) values and the influence of corneal parameters is established.
PURPOSE: This study compares IOP measurements made using different tonometers in patients implanted with ICRS and assesses the effects of central corneal thickness (CCT), corneal curvature, and corneal astigmatism on the IOP measurements obtained.
METHODS
In this cross-sectional study, IOP was measured using three different tonometers in 91 eyes of 91 patients with corneal ectasia implanted at least 6 months previously with ICRS. The tonometers tested were the TonoPen XL, Pascal dynamic contour tonometer (DCT), and iCare Pro rebound tonometer. GAT measurements were used as reference. Agreement among the IOPs provided by the different tonometers and the influence of corneal variables on the IOP measurements obtained were assessed using the Bland-Altman method, intraclass correlation coefficients, and multiple linear regression analysis.
RESULTS
Mean IOP differences were GAT versus TonoPen XL -0.8 ± 3.07 mm Hg, GAT versus DCT -1.0 ± 3.26 mm Hg, and GAT versus iCare Pro 0.8 ± 2.92 mm Hg. Our multiple linear regression analysis identified CCT as a confounding factor affecting all the tonometer readings but DCT-IOP.
CONCLUSIONS
In patients fitted with ICRS, IOP measurements made using the iCare Pro and TonoPen XL showed most agreement with GAT. Intraocular pressure measurements made by DCT were unaffected by corneal topographic factors though this procedure slightly overestimated GAT readings.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.