m+Medelevator.RU
← Поиск публикаций
PloS one · 2016

The Relationship between Corvis ST Tonometry and Ocular Response Analyzer Measurements in Eyes with Glaucoma.

Клиническое исследованиеСравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Matsuura Masato, Hirasawa Kazunori, Murata Hiroshi, Yanagisawa Mieko, Nakao Yoshitaka, Nakakura Shunsuke, Kiuchi Yoshiaki, Asaoka Ryo

Аннотация

It is important to compare the results of Corneal Visualization Scheimpflug Technology instrument (CST) measurements and Reichert Ocular Response Analyzer (ORA) parameters. The purpose of the study was to investigate the association between CST measurements and ORA parameters in ninety-five patients with primary open-angle glaucoma. Measurements of CST, ORA, axial length (AL), average corneal curvature (CC), central corneal thickness (CCT) and intraocular pressure (IOP) with Goldmann applanation tonometry (GAT) were carried out. The association between CST and ORA parameters was assessed using linear regression analysis, with model selection based on the second order bias corrected Akaike Information Criterion index. Measurements from ORA (corneal hysteresis [CH] and corneal response factor [CRF]) had high intraclass correlation coefficients (ICC) and low coefficients of variation, but some CST parameters showed much lower reproducibility, namely: A1 length, A2 length, highest concavity time and peak distance. Of 12 CST parameters tested, 8 were significantly correlated with CH and 10 were significantly correlated with CRF, however, the magnitude of the correlation coefficients were weak to moderate at best. The optimal model to explain CH using CST measurements was given by: CH = -76.3 + 4.6*A1 time + 1.9*A2 time + 3.1 * highest concavity deformation amplitude + 0.016*CCT (R2 = 0.67, p <0.001). Similarly, the optimal model for CRF was given by: CRF = -53.5 + 4.2*A1 time + 1.9*A1 length + 20.8*A1 deformation amplitude + 0.8*A2 time + 0.017*CCT (R2 = 0.73, p <0.001). ORA parameters show higher reproducibility than CST measurements. Although many CST parameters are significantly related to ORA parameters, the strengths of these relationships are weak to moderate.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты